摘要
介绍了一种自行研制的长程差分光学吸收光谱(DOAS)系统,该系统采用光电二极管阵列(PDA)作为光谱检测器,代替结构较为复杂的光电倍增管(PMT)加扫描圆盘的SD检测器,并对该系统检测器的偏置、暗电流、噪声、线性以及光谱仪的分辨率、光谱范围等性能进行了测试,同时对SO2,NO2等大气污染物进行了外场的测定,确定出相应光程下该系统测量SO2,NO2的最低检测限。
A long-path differential optical absorption spectroscopy (DOAS) system is introduced. A photodiode array is employed as the detector to replace the complicated SD detector which consists of a PMT and a slotted disk. The properties of the detector and the spectrometer unit such as offset, dark current, noise, linearity, resolution, and wavelength range were measured. This system was also tested to measure SO2 and NO2 in the atmosphere. The detection limits of this system for SO2, and NO2 over a 713 m light path were determined.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
2005年第9期1463-1467,共5页
Spectroscopy and Spectral Analysis
基金
国家高技术研究发展计划"863"(2002AA641020)
中国科学院安光所所长基金资助项目