期刊文献+

载气流速对高场不对称波形离子迁移谱的影响 被引量:24

Effect of Carrier Gas Flow Rate on High-Field Asymmetric Waveform Ion Mobility Spectrometry
下载PDF
导出
摘要 载气流速是影响高场不对称波形离子迁移谱(FAIMS)的重要参数。以自制的高场不对称波形离子迁移谱仪为实验平台,在射频电场幅值3kV/cm,频率500kHz,占空比0.36的条件下,研究了载气流速对苯离子迁移谱谱峰强度和半峰宽的影响。实验结果表明:载气流速为3.7L/min时,苯样品的谱峰强度最大,仪器的灵敏度最高。随着载气流速的增加,谱峰半峰宽变宽,仪器的分辨率下降。载气流速为3.0~3.7L/min时仪器综合性能最佳。此结果对于控制迁移谱仪载气流速有重要的参考意义。 Carrier gas flow rate is a key parameter that significantly affects the high-field asymmetric waveform ion mobility spectrometry.The effect of carrier gas flow rate on the benzene spectral peak intensity and peak width at half maximum has been investigated on the homemade high-field asymmetric waveform ion mobility spectrometer under the conditions of radio frequency electric field magnitude 30000 V/cm,frequency 500 kHz and dutyfactor 0.36.The experiment results showed that the flow rate at 3.67 L/min provides the maximum peak intensity and optimal sensitivity.The peak width at half maximum broadens and the resolving power decreased with the increase of carrier gas flow rate.The optimal spectrometer performance was obtained at the flow rates between 3 L/min and 3.67 L/h.The results and conclusions are useful for the carrier gas flow rate optimal control of high-field asymmetric waveform ion mobility spectrometry.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 2010年第7期1027-1030,共4页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金(No.60871037) 863计划(No.2007AA04Z341)资助项目
关键词 高场不对称波形离子迁移谱 载气流速 High-field asymmetric waveform ion mobility spectrometry Carrier gas flow rate
  • 相关文献

参考文献12

二级参考文献48

  • 1许峰,王海龙,关亚风.离子迁移谱研究进展[J].化学进展,2005,17(3):514-522. 被引量:40
  • 2袁曦,郑健,施亮,蒋大真.用于现场痕量检测的IMS爆炸物探测仪研究[J].核技术,2006,29(7):544-547. 被引量:16
  • 3时迎国,劭士勇,李安林,姚琏,王宾,李芳,王俊德,李海洋.迁移管的电场强度对真空紫外电离-离子迁移谱仪性能的影响[J].分析化学,2006,34(9):1353-1356. 被引量:13
  • 4COHEN M J,KARASEK F W. Plasma chromatogr-aphy - a new dimension for gas chromatography and mass spectrometry. J. Chromatogr. Sci., 1970,8(6):330-337.
  • 5COLLINS D C, LEE M L. Developments in ion mobility spectrometry-mass spectrometry. Anal. Bioanal.Chem., 2002,372(1):66-73.
  • 6http://www.oceta.on.ca/profiles/barringer/ionscan.html.
  • 7TAYLOR St J,PIPER L J, CONNOR J A,et al.Design aspects and operation characteristics of the lightweight chemical detector. IJIMS,1998,1(1):58-63.
  • 8HILL HERBERT H,SIMPSON GREG. Capabilities and limitations of ion mobility spectrometry for field screening applications.Field Anal. Chem. Technol., 1997,1(3):119-134.
  • 9KRYLOV E V. Comparison of the planar and coaxial field asymmetrical waveform ion mobility spectrometer (FAIMS). Int. J. Mass Spectrom., 2003,225(1):39-51.
  • 10EWING R G. ATKINSON D A, EICEMAN G A,et al. A critical review of ion mobility spectrometry for the detection of explosives and explosive related compounds. Talanta, 2001,54(3):515-529.

共引文献44

同被引文献227

引证文献24

二级引证文献76

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部