期刊文献+

2种肌红蛋白检测试剂的比较 被引量:1

下载PDF
导出
摘要 目的比较2种肌红蛋白检测试剂盒的敏感性和相关性。方法收集冠心病患者和健康体检者的血液标本,分别用肌红蛋白免疫比浊法检测试剂盒(北京九强生物技术股份有限公司)和肌红蛋白酶联荧光法检测试剂盒(法国生物梅里埃公司)进行肌红蛋白定量检测,对检测结果进行χ2检验和关联性分析。结果 120例冠心病患者标本中,九强试剂盒检出93例阳性(77.5%),梅里埃试剂盒检出62例阳性(51.7%),经χ2检验,2种试剂盒测定的阳性率差异有统计学意义(P<0.05)。60例健康体检者标本中,九强试剂盒检出1例阳性,梅里埃试剂盒未检出阳性,经χ2检验,2种试剂盒测定的阴性率差异无统计学意义(P>0.05)。经关联性分析,2种试剂盒的相关系数(r)=0.94。结论两种检测试剂盒检测结果关联性好,都具有很好的稳定性和特异性,但是九强试剂盒的敏感性明显优于梅里埃试剂盒,能够为医生提供更可靠的结果,有效降低临床漏诊率。
出处 《国际检验医学杂志》 CAS 2014年第2期229-230,共2页 International Journal of Laboratory Medicine
基金 国家高技术研究发展计划(863计划 2011AA02A111)
  • 相关文献

参考文献13

二级参考文献67

共引文献148

同被引文献23

  • 1Tunstall-Pedoe H, Kuulasmaa K, Amouyel P, et al. Myocardial infarction and coronary death in the World Health Organization MONICA Project-registration procedures, event rates, and casefatality rates in 38 populations from 21 countries in 4 continents[ J]. Circulation, 1994, 90 ( 1 ) : 583 - 612.
  • 2Kennedy RL, Harrison RF, Burton AM, et al. An artificial neural network system for diagnosis of acute myocardial infarction (AMI) in the accident and emergency department: evaluation and comparison with serum myoglobin measurements [ J ]. Comput Meth Prog Bio, 1997, 52(2) : 93 -103.
  • 3Mair J, Morandell D, Genser N, et al. Equivalent early sensitivities of myoglobin, creatine kinase MB mass, creatine kinase isoform ratios, and cardiac troponins I and T for acute myocardial infarction [J]. ClinChem, 1995, 41(9): 1266-1272.
  • 4Gilum RF, Fortmann SP, Prineas R J, et al. International diagnostic criteria for acute myocardial infarction and acute stroke [ J ]. Am Heart J, 1984, 108(1): 150-158.
  • 5Dara/n F, Yager P, Gan KL, et al. On-chip detection of myoglobin based on fluorescence [ J ]. Biosens Bioelectron, 2009, 24 ( 6 ) : 1744 - 1750.
  • 6Matveeva EG, Gryczynski Z, Lakowicz JR. Myoglobin immunoassay based on metal particle-enhanced fluorescence [ J ]. J Immunol Methods, 2005, 302 (1/2) : 26 -35.
  • 7Masson JF, Obando L, Beandoin S, et al. Sensitive and real-time fiber-optic-based surface plasmon resonance sensors for myoglobin and cardiac treponin I[J]. Talanta, 2004, 62(5) : 865 -870.
  • 8Yan L, Tang Y, Xiang J, et al. l~cn and application ~f monoclonal antibody against heavy trial chmnfium[ J]. Xi Bao Yu Fen Zi Mian Yi Xue Za Zhi, 2011,27(4) : 422 -42'$.
  • 9Raghava GP, Agrewala JN. Method for determining the affinity of monoclonal antibody using non-competitive ELISA: a computer program[J]. J lmmunoassay, 1994, 15(2): 115-128.
  • 10Collinson PO, Garrison L, Christenson RH. Cardiac biomarkers-A short biography[J]. Clin Biochem, 2015, 48(4/5) : 197 -200.

引证文献1

二级引证文献3