低浓度对比剂与低管电压在后门控CT冠脉成像中的可行性研究 被引量:7

Feasibility of Low Concentration Contrast Medium Combined with Low kV in Retrospectively Gated CT Coronary Angiography
下载PDF
导出
摘要 目的:比较低浓度对比剂与低管电压和常规高浓度对比剂与高管电压在后门控CT冠状动脉血管成像(CT coronary angiography,CTCA)中的图像质量.方法:2014年2月11日至2014年5月9日,对150名心率超过65次/min的患者行64排容积后门控CT冠状动脉血管成像扫描,根据他们的身体质量指数(BMI)分成3组(A组:n=50,BMI<23kg/m2,对比剂浓度为300mgI/ml,管电压为80kV;B组:n=50,23kg/m2≤BMI<28kg/m2,对比剂浓度为300mgI/ml,管电压为100kV;C组:n=50,BMI<28kg/m2,对比剂浓度为370mgI/ml,管电压为120kV),其中C组为常规剂量对照组.所有患者的对比剂的用量为70~80ml,注射速率为4.5~5ml/s.测量主动脉根部、左主干以及右主干的平均CT值,并根据下述的公式计算对比噪声比和信噪比.记录辐射剂量并且对图像质量进行评分.所有结果进行方差分析或者秩和检验.结果:三组患者主动脉根部、左主干以及右主干的对比噪声比和信噪比之间无统计学差异(P>0.05),三组患者所吸收的辐射剂量有显著统计学差异(P<0.05),低管电压组辐射剂量分别下降68%和44%,三组患者之间的图像质量无统计学差异(P>0.05).结论:采用低浓度对比剂与低管电压技术不会降低CTCA的图像质量,不影响诊断,可以运用于后门控CTCA. Purpose:To compare the image quality of low concentration contrast medium combined with low kV and high concentration contrast medium combined with 120 kV in retrospectively gated CT coronary angiography (CTCA).Methods:A total of 150 patients,each with a heart rate above 65bpm,underwent retrospectively gated CTCA on a 64-slice CT scanner.All the patients were divided into three groups according to (their) respective BMI(s) (group A:n=50,BMI〈23kg/m2,80kV,300mgI/ml; group B:n=50,23kg/m2≤BMI〈28 kg/m2,100kV,300mgI/ml; group C:n=50,BMI〈28kg/m2,120kV,370mgI/ml).The total volume of contrast medium was 70-80ml,and injection rate was 4.5-5.0ml/s.The CT values of aortic root(AR),left main artery(LMA),right main artery(RMA) and subcutaneous fat tissue were measured.The contrast-to-noise ratio (CNR) of AR,LMA and RMA were calculated.The values of CTDI and DLP were recorded.Image quality was assessed on 5-point scale.The results were compared between groups with one-way ANOVA.Results:The mean CNR and SNR of vessels in group A and group B was with no significant difference from group C(P〉0.05).The effective radiation dose in group A and group B was both lower than that in group C(P〈0.05),group A and group B fell by 68% and 44% in terms of effective radiation dose comparing to group C.There was no significantly difference on image quality scores between group A,group B and group C (P〉0.05).Conclusion:The use of the low concentration contrast medium and low tube voltage will not lower the image quality of CTCA and affect the diagnosis,which can be used in retrospectively gated CT coronary angiography.
出处 《中国医学计算机成像杂志》 CSCD 北大核心 2014年第5期446-450,共5页 Chinese Computed Medical Imaging
基金 上海市浦江人才 NO.14PJD028~~
关键词 对比剂 管电压 冠状动脉 血管成像 辐射剂量 Contrast medium Tube voltage Coronary Angiography Radiation dose
  • 相关文献

参考文献10

  • 1邵丹丹,林晓珠,刘燕,董海鹏,凌华威,陈克敏.心率对16层CT冠脉图像质量的影响[J].中国医学计算机成像杂志,2009,15(1):31-34. 被引量:2
  • 2王妍焱,吴国庚,周诚,高建华,焦晟,曹会志.64层螺旋CT前门控冠状动脉横断面扫描低剂量技术的初步研究[J].中华放射学杂志,2008,42(10):1018-1021. 被引量:79
  • 3Jung S C, Kim SH, Cho J Y. A Comparison of the use of contrast media with different iodine concentrations for multidetector CT of the kidney. Korean J Radiol, 2011, 12: 714-721.
  • 4Nakaura T, Kidoh M, Sakaino N, et al. Low contrast- and love radiation dose protocol for cardiac CT of thin adults at 256-row CT: usefulness of low tube voltage scans and the hybrid iterative reconstruction algorithm. Int J Cardiovasc Imaging, 2013, 29: 913-923.
  • 5Sommer WH, Schenzle JC, Becker CR, et al. Saving dose in triple- rule-out computed tomography examination using a high-pitch dual spiral technique. Invest Radiol, 2010, 45: 64-71.
  • 6Leschka S, Kim CH, Baumueller S, et al. Scan length adjustment of CT coronary angiography using the calcium scoring scan; effect on radiation dose. AJR, 2010, 194: 272-277.
  • 7Sany, D, Refaat, H, Elshahawy, Y, et al. Frequency and risk factors of contrast-induced nephropathy after cardiac catheterization in type II diabetic patients: a study among Egyptian patients. Ren Fail, 2014, 36: 191-197.
  • 8Paul J-F. Individually adapted coronary 64-slice CT angiography based on precontrast attenuation values, using different kVp and tube current settings: evaluation of image quality. Int J Cardiovasc Imaging, 2011, 27.. 53-59.
  • 9Abada HT, Larchez C, Daoud B, et al. MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. A JR, 2006, 186(6 Suppl 2): 387-390.
  • 10张兆琪,徐磊.重视冠状动脉多层CT成像的低剂量检查[J].中华放射学杂志,2009,43(7):681-683. 被引量:80

二级参考文献24

  • 1张森,杜湘珂,李剑颖.心电图调制电流对于容积CT心脏扫描图像质量及曝光剂量影响的评估[J].中华放射学杂志,2006,40(9):974-976. 被引量:45
  • 2Trabold T, Buchgeister M, Kuttner A, et al. Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo, 2003, 175 : 1051-1055.
  • 3Mallet NR, Cademartiri F, van Mieghem CA, et al. High- resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation, 2005,112 : 2318-2323.
  • 4Flohr TG, Schoepf U J, Kuettner A, et al. Advances in cardiac imaging with 16-section CT systems. Acad Radiol, 2003, 10: 386-401.
  • 5Hsieh J, Londt J, Vass M, et al. Step-and-shoot data acquisition and reconstruction for cardiac X-ray computed tomography. Med Phys, 2006,33:4236-4248.
  • 6Hoffmann MH, Shi H, Manzke R, et al. Noninvasive coronary angiography with 16 - detector row CT: effect of heart rate. Radiology, 2005, 234:86 - 97.
  • 7Nieman K, Rensing B J, van Geuns R J, et al. Non - invasive coronary angiography with multislice spiral computed tomography: impact of heart rate. Heart, 2002, 88:470-474.
  • 8Austen WG, Edwards JE, Frye RL, et al. Areporting system on patients evaluated for coronary artery disease: report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation, 1975,51: 5- 40.
  • 9Schroeder S , Kopp AF , Kuettner A , et al . Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography: experience in 94 patients. Clin Imaging, 2002. 26 : 106- 111.
  • 10Hong C, Becker CR, Huber A, et al. ECG- gated reconstructed multi - detector row CT coronary angiography : effect of varying trigger delay on image quality. Radiology, 2001,220: 712- 717.

共引文献151

同被引文献47

引证文献7

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部