期刊文献+

脉冲回波法与水听器法测量超声探头声场参数比较研究 被引量:6

Experimental Research of Push-echo and Hydrophone Methods for Measurement of Sound Field of Ultrasonic Transducers
下载PDF
导出
摘要 针对美国ASTM E1065中提及的脉冲回波法与水听器法,通过在〈10 MHz频率范围内的实验,比较了这两种方法在评价平面超声探头与聚焦超声探头声场参数的一致性。实验表明,针对超声探头辐射声场参数的测量,在合理选择小球参数的基础上采用脉冲回波法与水听器法具有较好的测量一致性,而水听器法则提供了一种具有量值可溯源的超声探头评价与校准方法。 When the frequency range of ultrasonic transducers is less 10 MHz, experimental researches, used thepulse-echo and hydrophone methods introduced in the ASTM E1065, has compared the consistency of two methods formeasurement of sound field parameters of the plane ultrasonic transducers and the focused transducers. Used the reasonablediameters of the small balls, the experiments show well consistency of pulse-echo and hydrophone methods for measurementof sound field parameters of ultrasonic transducers, but the hydrophone method is a method of measurement traceability inevaluation and calibration of ultrasonic transducers.
作者 韩冥生 杨平
出处 《计量学报》 CSCD 北大核心 2015年第2期166-170,共5页 Acta Metrologica Sinica
基金 国家科技支撑计划课题(2011BAI02B03)
关键词 计量学 超声探头 声场参数 脉冲回波 水听器 量值溯源 Metrology Uhrasonie transducer Sound field parameter Pulse-echo Hydrophone Measurementtraceability
  • 相关文献

参考文献11

  • 1国家机械工业局.JB/T10062-1999超声探伤用探头性能测试方法[S].1999.
  • 2国家质检总局.GB/T18694-2002超声检验探头及其声场表征[S].2002.
  • 3ASTM. ASTM E1065 -08 Standard Guide for Evaluating Characteristics of Ultrasonic Search Units [ S ]. 2008.
  • 4BS. BS EN12668-2:2010 Non-destructive testing-- Characterization and verification of ultrasonic examination equipment[ S]. 2010.
  • 5Lucas B G, Muir T G. The field of a focusing source[ J]. J Acoust Soc Am, 1982, 72(4) :1289 - 1296.
  • 6Xucai Chen, Schwarz K Q, Parker K J. Radiation pattern of a focused transducer: A numerically convergent solution[ J ] . J Acoust Soc Am, 1993, 95 ( 5 ) : 2979 - 2991.
  • 7IEC. IEC 62127-3 Ultrasonics-hydrophone - Part3 : Properties of hydrophones for ultrasonic fields up to 40 MHz[ S]. 2007.
  • 8Zeqiri B, Hodnett M, Smith R A, et al. An assessment of pulse-echo and hydrophone methods for characterising ultrasonic NDT transducers [ J ] . Insight-Non-Destructive Testing and Condition Monitoring, 1998, 40 ( l ) : 20 - 27.
  • 9Zeqiri B, Hodnett M. Standardisation of procedures for ultrasonic inspection of polymer matrix composites : transducer characterisation and materials evaluation [ R ]. NPL Report CMAM 9, 1997.
  • 10Yanqiang Feng, Haijiang Zhu, Ping Yang. Visualization and measurement of the acoustic field of the piston transducer[J]. Sensor Review, 2013, 33(2): 174- 180.