摘要
建立采用离子色谱法测定复方电解质眼内冲洗液中钠、钾、镁、钙离子含量的分析方法。以20 mmol/L甲烷磺酸作淋洗液,流速1.0 mL/min,柱温30℃,25μL样品经Ion Pac TM CG12A(4 mm×50 mm)阳离子保护柱与Ion Pac TM CS 12A(4 mm×250 mm)阳离子交换柱后,利用电导检测器,采用标准曲线法测定样品溶液中各阳离子的含量。当钠离子浓度范围为47.20~93.08μg/mL,钾离子浓度范围为5.42~10.14μg/mL,镁离子浓度范围为0.50~0.93μg/mL,钙离子浓度范围为1.92~3.56μg/mL时,线性关系良好,相关系数(r)均大于0.99,各阳离子的加标回收率均处于99%~102%(n=6),同时相对标准偏差均低于1.0%(n=6),该方法测定结果具有较好的准确度和精密度,可推广用于相关药品的质量控制研究。
To establish a method for the content simultaneous determination of sodium ion,potassium ion,magnesium ion and calcium ion in compound electrolyte intraocular irrigating solution by ion chromatography.The mobile phase was 20 mmol/L methane sulfonic acid solution,the flow rate was 1.0 mL/min,the temperature of column was 30 ℃,the injection volume was 25 μL,and chromatographic separation was performed on a Dionex Ion Pac TM CG12A guard column(4 mm×50 mm)and Dionex Ion Pac TM CS12A cation column(4 mm×250 mm),electrical conductivity detector was adopted and the results were obtained by standard curve of different cations.The Na ^+,K ^+,Mg ^2+ and Ca ^2+ had good linear relationship in the ranges of 47.20~93.08,5.42~10.14,0.50~0.93 and 1.92~3.56 μg/mL,respectively.All of the correlation coefficients were more than 0.99.The recoveries for standard markers were added were in the range of 99%~102%,and RSD less than 1.0%.Therefore,the results of new method have preferable accuracy and precision,thus could be used in quality control of this drug.
作者
张晓明
ZHANG Xiao-ming(Chemical Engineering and New Energy Material College,Zhuhai College of Jilin University,Zhuhai 519041,China)
出处
《应用化工》
CAS
CSCD
北大核心
2019年第3期736-739,共4页
Applied Chemical Industry
基金
2017广东省教育厅重点平台和重大项目(2017KTSCX212)
2017吉林大学珠海学院创新能力培育工程项目(2017XJCQSQ031)
吉林大学珠海学院"三个层次"人才建设工程项目资助
关键词
离子色谱
复方电解质眼内冲洗液
阳离子
含量测定
ion chromatography
compound electrolyte intraocular irrigating solution
cation
content determination