期刊文献+

不确定度评价方法在高纯锗探测器参数修正中的应用 被引量:2

Application of uncertainty evaluation method in parameter correction of high-purity germanium detector
下载PDF
导出
摘要 将统计性不确定度评价方法应用到高纯锗(HPGe)探测器参数的标定和修正工作中.结合蒙特卡洛模拟和统计性不确定度评价方法,分析探测器参数在探测效率模拟结果中的相对重要性,对重要参数同时进行多次简单蒙卡抽样,模拟计算探测器对多空间点、多γ源的探测效率,找出真实效率与模拟效率的最小偏差,即能确定最佳探测器参数.结果表明,此方法修正探测器参数后,对60Co,137Cs,241Am源在标定点的模拟效率与真实效率相对误差均小于0.5%,在3个验证点的模拟效率与真实效率相对误差均小于0.5%. In view of the defect of commonly used method for correcting the parameters of high-purity germanium(HPGe)detector,a new correction method based on statistical uncertainty analysis and Monte Carlo simulation is proposed.The relative importance of the parameters of HPGe detector in the simulation results is analyzed,to find important parameters.The simple Monte Carlo sampling of important parameters is conducted thousands times,from which the sampled values are used as the detector model\s parameters.Then the detection efficiencies of the detector for multipleγsources located at different spatial points are simulated.The best parameters of detector can be determined by the minimum deviation between true efficiencies and simulation efficiencies.Take these best parameters as the detector model’s parameters,the results show that the relative errors between the simulation efficiencies and the true efficiencies for the sources 60Co,137Cs,241Am at the calibration points are all less than 0.5%,the relative errors between the simulation efficiencies and the true efficiencies at 3 verification points are all less than 0.5%.
作者 李自维 白立新 张一云 LI Zi-Wei;BAI Li-Xin;ZHANG Yi-Yun(College of Physics,Sichuan University,Chengdu 610065,China)
出处 《四川大学学报(自然科学版)》 CAS CSCD 北大核心 2020年第5期961-967,共7页 Journal of Sichuan University(Natural Science Edition)
关键词 不确定度评价方法 高纯锗探测器 探测效率 敏感性分析 Uncertainty evaluation method High-purity germanium detector Detection efficiency Sensitivity analysis
  • 相关文献

参考文献7

二级参考文献32

  • 1向长兴.一组实用γ谱分析效率刻度曲线[J].核电子学与探测技术,1994,14(6):363-366. 被引量:9
  • 2张斌全,马吉增,程建平,刘立业,毛永.蒙特卡罗方法计算用于低能光子测量的高纯锗探测器的效率[J].核电子学与探测技术,2005,25(3):274-277. 被引量:26
  • 3向东,郭兰英.γ能谱的蒙特卡罗模拟[J].数理医药学杂志,2006,19(3):229-231. 被引量:16
  • 4丁爱平,李莹,卢磊,曾勤,郑善良,吴宜灿.粒子输运计算模型M CNP模型的可视化实现[J].原子核物理评论,2006,23(2):130-133. 被引量:12
  • 5Alonso M,et al. Monte Carlo estimation of absorbed dose to organs in computed tomography[J].Health Phys,2002,82(2):233.
  • 6Stabin MG,et al.Photon specific absorbed fractions culculated in the trunk of an adult male voxel-based phantom[J]. Health Phys,2002,82(1):21.
  • 7Borisov N,. et al.A new graphical user interface for fast construction of computation phantoms and mcnp calculations:application to calibration of in vivo measurement systems[J]. Health Phys,2002,83(2):272.
  • 8Bozkurt A, et al Fluence-to-dose conversion coefficients from monoenergetic neutrons below 20 MeV based on the VIP-Man anatomical model[J]. Phys Med Biol,2000,(45):3059.
  • 9Moens L, et al Calculation of the absolute peak efficiency of gamma-ray detectors for different counting geometries[J]. Nucl Instrum Methods,1981,187:451.
  • 10Moens L, et al Calculation of the peak efficiency of high-purity Germanium detectors[J]. Int J Appl Radiat Isot,34(8):1085.