基于ZYNQ的ICP-MS数据采集处理系统设计与实现 被引量:4

Design and Implementation of ICP-MS Data Acquisition and Processing System Based on ZYNQ
下载PDF
导出
摘要 数据采集系统是电感耦合等离子体质谱仪(ICP-MS)的重要组成部分。本研究针对ICP-MS需要同步采集脉冲和模拟信号,且信号信噪比低的问题,提出了一种基于ZYNQ芯片的数据采集处理解决方案。利用ZYNQ集处理器和可编程逻辑于一体的特性,在ARM端搭载了用于流程控制及以太网通讯的FreeRTOS实时操作系统,在FPGA端基于FIR compiler和MCDMA IP核实现了数据的采集、缓存、动态滤波及传输。脱机测试结果表明,脉冲计数误差率小于0.0005%,模拟计数线性度高达0.99998,数字滤波幅频特性曲线与理论基本一致。在本团队自行研制的ICP-MS仪器中,利用本系统测试标准调谐液TUNE D,数据采集满足仪器需求,配置64阶低通滤波后,模拟计数中^(115)In谱峰信噪比提高了49.59%。该系统FPGA端各资源占用率均小于50%,片上总功耗仅为2.28 W,有利于推动质谱仪小型化、智能化发展。 Data acquisition system is an important part of inductively coupled plasma mass spectrometer(ICP-MS).Aiming at the problem that ICP-MS pulse and analog signal need to be collected synchronously and the signal-to-noise ratio is low,a data acquisition and processing solution based on ZYNQ chip were proposed.Using the characteristics of ZYNQ integrating processor and programmable logic,the FreeRTOS real-time operating system for process control and Ethernet communication was equipped on the ARM side.Three modules were designed on the FPGA side to realize data acquisition,cache,dynamic filtering and transmission,including counting module for synchronous acquisition of ICP-MS pulse signal and analog signal,FIR filter module supporting online configuration based on FIR compiler IP core,and dual channel data transmission module based on MCDMA IP core.The dynamic configuration of FIR filter module was controlled by ARM,and the filter coefficients were generated by ARM calculation to improve the universality of the system.Moreover,the communication between ARM and FPGA and between modules were all based on AXI4 bus.The Modbus/TCP transport protocol was used for Ethernet communication to ensure the stability and universality of the system.The system test was divided into two steps of offline test and online test.The offline test verified the function and performance of each module in the system.The performance of the pulse count module was tested by using the pulse signal generated by the signal generator AFG320 to simulate ICP-MS.After eliminating the error caused by the crystal vibration deviation,the relative error of the pulse count was less than 0.0005%.The ADC in the analog count was calibrated by using DP832 DC regulator power supply and 34401A six-bit half-digital multimeter.The result showed that the fitting linearity is as high as 0.99998.The digital filter performance was tested by inputting sinusoidal signals of different frequencies,and the results showed that the amplitude-frequency characteristic curve is basically consistent with the theory.A standard tuning solution,tune D(Thermo Fisher Scientific),was tested in our team's self-study ICP-MS instrument in-house using this system.It was found that the pulse count had little noise,while the analog count had a large amount of noise thereby affecting the subsequent spectral peak feature extraction.The noise spectrum distribution in the data was analyzed by FFT analysis of the analog count results,and the appropriate parameters of FIR digital filter were selected according to the characteristics of Gaussian peaks.After filtering,the analog count spectrogram is improved significantly,the main noise intensity attenuation is greater than 40 dB,and the signal-to-noise ratio of ^(115)In peak is increased by 49.59%.The filtered spectrum fully meets the requirements of subsequent peak feature extraction.The resource occupancy rate of the FPGA end of the system is less than 50%,and the total power consumption on the chip is only 2.28 W,which is conducive to promoting the miniaturization and intellectualization of mass spectrometry.In the future,algorithms can be designed at the ARM side or upper computer end to automatically adjust the filter parameters and further improve the degree of system automation.
作者 苏伟童 韩文念 王康 田霖 赵珍阳 张勇 田中朝 汪曣 SU Wei-tong;HAN Wen-nian;WANG Kang;TIAN Lin;ZHAO Zhen-yang;ZHANG Yong;TIAN Zhong-chao;WANG Yan(Tianjin University,Tianjin 300072,China;Tianjin AIMS Instrument Co.,Ltd,Tianjin 300134,China;Shandong Dongyi Photoelectric Instrument Co.,Ltd.,Yantai 264000,China)
出处 《质谱学报》 EI CAS CSCD 北大核心 2022年第4期473-481,I0004,共10页 Journal of Chinese Mass Spectrometry Society
基金 山东省重大科技创新工程项目“三重四极杆电感耦合等离子体质谱仪的研制及产业化”(2018YFGH0302)。
关键词 电感耦合等离子体质谱(ICP-MS) ZYNQ 数据采集 动态数字滤波 inductively coupled plasma mass spectrometry(ICP-MS) ZYNQ data acquisition dynamic digital filtering
  • 相关文献

参考文献3

二级参考文献17

  • 1Quistgaard JU. Signal acquisition and processing in medical diagnostic ultrasound [ J ]. IEEE Signal Processing Magazine, 1997,14:67 - 73.
  • 2冯若,刘忠齐,姚锦钟,等.超声诊断设备原理及设计[M].北京:中国医药科技出版社,1991.
  • 3Christopher JH, James MP, Robert JE, et al. Advance in ultrasound [J].Clinical Radiology, 2002, 57 : 157 - 177.
  • 4Patrick L, Ah M. Area-efficient FIR filter design on FPGA using distributed arithmetic [ A ]. In: IEEE International Symposium on Signal Processing and Information Technology [ C ]. Vancouver: IEEE,2006,6:248 - 252.
  • 5Girard P, Hbron O, Pravossoudovitch S, et al. Delay fault testing of look-up tables in SBAM based FPGA [J]. Journal of Electronic Testing, 2005, 21( 1 ) : 43 - 55.
  • 6EDA先锋工作室.AheraFPGA-CPLD设计[M].北京:人民邮电出版社.2005.
  • 7Abed KH, Vivek V, Nerurkar SB. High speed digital filter design using minimal signed digital representation [A]. In : The IEEE Southeast Conference[C]. Lauderdale: IEEE, 2005. 105 -110.
  • 8Uwe MB. Digital Signal Processing with Field Programmable Gate Arrays[M]. Berlin: Springer-Verlag, 2002.
  • 9Punchalard R, Lorsawatsiri A, Loetwassana W, ctal. Direct frequency estimation based adaptive algorithm for a second-order adaptive FIR notch filter[ J ]. Signal Processing, 2008,88 ( 2 ) : 315 - 325.
  • 10Labrosse Jean J.嵌入式实时操作系统μC/OS-Ⅱ.第2版.邵贝贝译.北京:北京航空航天大学出版社,2003

共引文献40

同被引文献26

引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部