摘要 目的分析Leksell Icon型伽玛刀锥形束CT的常规质量控制结果,评估其长期稳定性。方法使用QA TOOL plus验证锥形束CT精度;扫描Catphan 503模体,分析图像空间分辨率、对比度-噪声比和均匀性。结果锥形束CT图像体积的最大偏差为0.09~0.17 mm,均通过精度测试;预设的CT剂量指数2.5和6.3 mGy扫描模式下,空间分辨率很稳定,分别为7和8 lp/cm;对比度-噪声比和均匀性均符合参考要求。结论12个月的Leksell Icon型伽玛刀锥形束CT的常规质量控制比较稳定。锥形束CT的质量控制除了参考厂商的调试基线值外,针对伽玛刀统一的分析评价标准有待进一步完善。 Objective To analyze the conventional quality control result of Leksell Icon Gamma Knife cone-beam CT and evaluate long-term stability of cone-bem CT.Methods QA TOOL Plus was used to verify the accuracy of cone-beam CT.The phantom Catphan 503 was scanned,and the image spatial resolution,contrast to noise ratio and homogeneity were analyzed.Results The maximum deviation in image volume of cone-beam CT was 0.09-0.17 mm,which passed the accuracy test.At the scanning patterns with CT dose index of 2.5 and 6.3 mGy,the spatial resolution was very stable at 7 and 8 lp/cm respectively.The contrast noise ratio and uniformity meet the reference requirements.Conclusions The conventional quality control results of Leksell Icon Gamma Knife cone-beam CT are stable in 12 months.In addition to referring to the manufacturer′s baseline value,the unified analysis and evaluation standard for Gamma Knife need to be further improved for the quality control of cone-beam CT.
作者 蔡俊涛 徐云波 李鹏 Cai Juntao;Xu Yunbo;Li Peng(Department of Neurosurgery,West China Hospital,Sichuan University,Chengdu 610041,China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2023年第1期63-67,共5页 Chinese Journal of Radiological Medicine and Protection
关键词 伽玛刀 锥形束CT 质量控制 图像质量 Gamma knife Cone-beam CT Quality control Image quality
  • 相关文献

参考文献6

二级参考文献35

共引文献34

同被引文献16

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部