摘要 AIM:To invastigate intra-and interobserver reliability of interferometry,tear meniscus height(TMH)measurement and meibography(MBG)of an ocular surface analyzer,LacryDiag(Quantel Medical,France).METHODS:Five consecutive measurements and subsequent analysis of interferometry,TMH,and MBG were recorded by two examiners using the LacryDiag.To assess intra-and interobserver reliability,we used Cohen’s kappa for categorical variables(interferometry),or intraclass correlation coefficient for continuous variables(TMH,MBG).RESULTS:Thirty eyes of 30 examinees were included.For both observers,there was excellent intraobserver reliability for MBG(0.955 and 0.970 for observer 1 and 2,respectively).Intraobserver reliability for observer 1 was substantial for interferometry(0.799),and excellent for TMH(0.863).Reliability for observer 2 was moderate for interferometry(0.535)and fair to good for TMH(0.431).Interobserver reliability was poor for interferometry(0.074)and fair to good for TMH(0.680)and MBG(0.414).CONCLUSION:LacryDiag ocular surface analyzer in our study proves to be a reliable noninvasive tool for the evaluation of TMH and MBG.As for interferometry,poor interobserver reliability,fair to good intraobserver reliability for observer 1,and moderate for observer 2,leave room for improvement.