期刊文献+

流动注射在线分离富集火焰原子吸收法快速顺序测定环境样品中六种元素 被引量:4

Fast Sequential Determination of Cu,Pb,Cd,Co,Ni and Mn in Environmental Samples by FI-FAAS with On-line Separation and Enrichment
下载PDF
导出
摘要 本文采用双柱富集的在线分离富集系统与 2 2 0FS顺序多元素原子吸收分光光度计联用 ,以PT C18色谱预处理柱为预富集柱 ,二乙基二硫代氨基甲酸钠 (DDTC)和吡咯啶二硫代氨基甲酸铵 (APDC)混合物为螯合剂 ,甲醇为洗脱剂 ,首次实现了Cu ,Pb,Cd ,Co ,Ni,Mn六种元素的快速同时测定。富集倍数为 10 2~ 2 1 7之间。34s富集 (4 4 0mL)的检出限分别为Cu 2 91,Pb 5 81,Cd 0 35 ,Co 3 6 1,Ni4 0 5和Mn 1 94 μg·L-1,相对标准偏差分别为Cu 2 6 4 % ,Pb 2 93% ,Cd 2 5 9% ,Co 2 5 7% ,Ni 2 78%和Mn 3 6 8%。该方法操作简便、快速 ,用于环境样品中痕量元素的测定 ,得到了满意的结果。 Six elements, namely Cu, Pb, Cd, Co, Ni and Mn, were determined simultaneously and rapidly by using a double per-columns on-line enrichment system together with a 220FS sequential multi-elemental atomic absorption spectrophotometer. PT-C-18 column was used as pre-enrichment column, the mixture of DDTC and APDC were used as a chelating agent, and methanol was used as an eluent. The enrichment factors between 10.2 - 21.7 for the elements were obtained. The detection limits obtained with 34 seconds of enrichment (4.40 mL) for Cu, Pb, Cd, Co, Ni and Mn were 2.91, 5.81, 0.35, 3.61, 4.05 and 1.94 mug(.)L(-1), respectively. The relative standard deviation for Cu, Pb, Cd, Co, Ni and Mn were 2.64%, 2.93%, 2.59%, 2.57%, 2.78% and 3.68%, respectively. The proposed method was rapid and simple, and satisfactory results were obtained for trace element determinations in environmental samples.
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 2003年第4期785-788,共4页 Spectroscopy and Spectral Analysis
关键词 流动注射 在线分离富集 火焰原子吸收法 顺序测定 环境样品 痕量元素 flow injection analysis atomic absorption spectrometry PT-C-18 pre-column copper lead cadmium cobalt nickel manganese
  • 引文网络
  • 相关文献

参考文献5

二级参考文献22

共引文献58

同被引文献434

引证文献4

二级引证文献31

;
使用帮助 返回顶部