Sysmex XE-2100检测网织红细胞计数评价 被引量:4

Evaluation of sysmex XE-2100 in counting reticulocytes
下载PDF
导出
摘要 目的对SysmexXE-2100型全自动血细胞分析仪定量检测网织红细胞(RC)的性能进行评价。方法采用SysmexXE-2100定量检测RC,并与手工法进行比较。结果批内平均变异系数(CV)3.69%,天间CV4.36%,携带污染率为零;RC绝对值在(4.0 ̄770.0)×109/L范围内线性良好(r=0.998);采用SysmexXE-2100检测78例标本的RC值与手工法呈显著正相关(r=0.977,P<0.001)且结果差异无显著性(P>0.05)。结论SysmexXE-2100检测RC准确性好、精密度高,操作简便快速,并能提供较多的RC相关参数,可广泛应用于临床。 [Objective]To evaluate the effect of automated Reticulocyte counting by Sysmex XE-2100. The blood samples of patients and normal subjects were detected by XE-2100 and Miller's method recommended by International Committee for standardization in Haematology(ICSH). The linearity of Sysmex XE-2100 was perfect when RC counting was (4.0~770.0)×109/L (r =0.998), with intra-assay CV 3.69%and inter-assay CV 4.36%, carryover rate of 0%; There were highly correlation between Sysmex XE-2100 and Miller's method in 78 results of patients and normal subjects (r =0.977, P <0.001), and the results were no significant differences (P >0.05). [Conclusion] The RC counting by Sysmex XE-2100 is a highly accurate, precision and easily operated method, it also can provide many new parameters for clinical usage.
出处 《中国现代医学杂志》 CAS CSCD 北大核心 2005年第9期1404-1406,共3页 China Journal of Modern Medicine
关键词 网织红细胞计 SYSMEX XE-2100流式细胞术 贫血 reticulocyte counting sysmex XE-2100 flow cytometry anemia
  • 相关文献

参考文献8

二级参考文献15

  • 1丛玉隆,贾淑芳,王淑娟.网织红细胞计数法进展及质量控制[J].中华医学检验杂志,1994,17(6):377-380. 被引量:12
  • 2李百泉,国外医学临床生物化学与检验学分册,1991年,12卷,186页
  • 3张月香,国外医学临床生物化学与检验学分册,1991年,12卷,183页
  • 4王淑娟,实验诊断学,1991年
  • 5戚仁铎,诊断学(第3版),1990年
  • 6李影林,临床医学检验手册,1987年
  • 7王振生,实验检查的临床分析与诊断,1986年
  • 8叶应妩,新编临床检验法,1986年
  • 9周振业,临床检验杂志,1984年,2卷,34页
  • 10丛玉隆,上海医学检验杂志,1996年,11卷,7页

共引文献65

同被引文献28

引证文献4

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部