摘要
建立了氟的铝-钙指示剂络合物荧光猝灭测定法。本法λ=570nm,λ_(em)=600nm,线性范围为5~160ng/mLF ̄-。对5ng/mIF ̄-重复11次测定结果的相对标准偏差为2.4%,可直接用于温泉水中氟的测定。
A fluoro-quenching method for the de-termination of fluoride with Al-calconcar-boxylic acid complex is established.Theλ_(ex) andλ_(em) are 570nm and 600nm。The linearrange is 5~160ng/mL F ̄-, DL=3.8ng/mL。The RSD is 2.4%(11 measurements of 5ng/mL F ̄-).The method has been directly usedto determine fluoride in hot spring water.
出处
《分析试验室》
CAS
CSCD
北大核心
1995年第3期30-32,共3页
Chinese Journal of Analysis Laboratory
关键词
荧光猝天法
氟
钙指示剂
温泉水
Fluoro-quenching method
fluoride determination
calconcarboxylicacid.