期刊文献+

Sysmex XE-2100全自动血细胞分析仪对形态异常细胞提示功能评价 被引量:12

Clinical evaluation of the Sysmex XE-2100 analyzer in application of the white blood cells classification
下载PDF
导出
摘要 目的对Sysmex XE-2100全自动血细胞分析仪白细胞分类及形态异常细胞提示功能进行评价。方法对1200例各种住院病人标本,进行血片复检,评价其对异常细胞报警系统的性能。结果Sysmex XE-2100全自动血细胞分析仪对形态异常细胞提示系统具有很高灵敏度,其异常情况报警系统的灵敏度为100%,阴性预示值为100%,阳性预示值为79.55%,特异性是95.60%,可疑警号的假阳性率为12.83%,总检出有效率为96.25%。结论Sysmex XE-2100全自动血细胞分析仪对白细胞分类及形态异常细胞报警系统的功能良好,特异性好,灵敏度高,符合临床应用要求。 Objective To appraise the performance of the Sysmex XE-2100 analyzer to analyzer the white blood cells and its abnormal cells. Methods In order to research its application in analyzing of the abnormal white blood cells and alarms capability of detection abnormality, 1 200 samples were examined repeatedly by microscope measurement and a statistical comparison was also made. Results The Sysmex XE-2100 analyzer has a higher precision and the sensitivity of the alarming system was 100%. The predictive value of negative was 100%. The predictive value of positive was 79.55%. The specificity was 95.60%. The false positive rate was 12.83%. the efficiency rate was 96.25%. Conclusions The Sysmex XE-2100 analyzer has a good speciality, veracity and sensitivity in analyzing blood white cells and the abnormal cells. It can be used in the clinical laboratory.
出处 《实用全科医学》 2007年第3期257-258,共2页 Applied Journal Of General Practice
关键词 形态异常细胞 XE-2100血细胞分析仪 报警系统 评价 Abnormal cell-form Sysmex XE-2100 analyzer The alarm system Evaluate
  • 相关文献

参考文献2

  • 1R Herklote,AR Huber.Precision and accuracy of the leukocyte differ-ential on the XE-2100[].Sysmex J Int.2001
  • 2H Inoue.Overview of automated hematology analyzer XE-2100[].Sysmex J Int.1999

同被引文献70

引证文献12

二级引证文献52

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部