Hemolysis Test on a Magnetic Suspension Centrifugal Ventricular Assist Device
下载PDF
导出
摘要 通过体外以及在体溶血实验,探讨磁悬浮驱动离心式心室辅助装置血液相容性能.将左心室辅助装置置于充满新鲜羊血的模拟循环通路中,在输出流量为5 L/min、后负荷压力为100 mmHg的条件下,检测血泵体外溶血性能.在8只健康绵羊体内植入血泵进行在体辅助实验,通过定期抽血检验血浆游离血红蛋白值,测定其在体溶血性能.结果表明:可植入磁悬浮驱动离心式左心室辅助装置的体外标准溶血指数(normal index of hemolysis,NIH)为(0.075±0.017)mg/L.在体实验中,2例存活超过20 d,循环辅助后实验羊血浆游离血红蛋白(free hemoglobin,FHB)呈现上升趋势,2例实验FHB最大值分别达0.846和1.09 g/L,之后逐渐下降到术前水平.本实验血泵同其他国产血泵比较,具有较好的体外和在体血液相容性,具备好的长期动物在体辅助实验基础. To explore the hemolytic characteristics of a self-made implantable magnetic suspension centrifugal ventricular assist device in vitro and in vivo. The blood pump of the ventricular assist device was connected in a mock circulatory loop, filled with fresh sheep blood, and the pump generated 5 L/min flow against a head pressure of 100 mmHg. The vitro hemolytic properties were evaluated by determining the level of plasma free hemoglobin (FHB) at defined intervals. Blood pump was implanted in two healthy sheep in vivo test. After pumping, FHB and hemoglobin (HB) were detected continuously during the experiments. The normal index of hemolysis (NIH) in vitro was (0. 075 ±0. 017) mg/L. Vivo experiment results showed that the levels of FHB gradually increased, and the maximal mean values were 0. 846 and 1.09 g/L, respectively. Compared with others domestic LVADs, the vitro and vivo hemolyticproperties of this blood pump are satisfactory. Therefore, long-term circulation assist experiment of the new blood pump can be carried out to evaluate its hemodynamic and end-organ effect in the future.
出处 《北京工业大学学报》 CAS CSCD 北大核心 2013年第10期1596-1600,共5页 Journal of Beijing University of Technology
基金 国家863计划资助项目(2012AA041605) 首都卫生行业科研专项资助项目(2011-1005-02) 北京市自然科学基金资助项目(3112011)
关键词 心室辅助装置 溶血实验 动物实验 ventricular assist device hemolysis test animal tests
  • 相关文献

参考文献15

二级参考文献22

  • 1蔺嫦燕,侯晓彤,吴广辉,李冰一,王景,潘红九.XZ-Ⅱ型轴流血泵的流场分析[J].北京生物医学工程,2005,24(6):405-409. 被引量:11
  • 2张岩,薛嵩,桂幸民,胡盛寿,孙寒松,朱晓东.运用三维数值模拟对人工心脏轴流血泵的设计和改进[J].中国生物医学工程学报,2007,26(1):35-41. 被引量:20
  • 3罗新锦,孙寒松,李楠,胡盛寿,许建屏,张岩.机械循环辅助对术后患者血浆游离血红蛋白的影响[J].中国胸心血管外科临床杂志,2007,14(3):173-176. 被引量:6
  • 4Chinn R, Dembitsky W, Eaton L, et al. Multicenter experience: prevention and management of left ventricular assist device infections. ASAIOJ, 2005, 51(4):461-470.
  • 5Haddad H, Elabbassi W, Moustafa S, et al. Left ventricular assist devices as bridge to heart transplantation in congestive heart failure with pulmonary hypertension. ASAIO J, 2005, 51 (4) :456-460.
  • 6Davies RR, Russo MJ, Hong KN, et al. The use of mechanical circulatory support as a bridge to transplantation in pediatric patients:an analysis of the United Network for Organ Sharing database. J ThoracCardiovasc Surg, 2008, 135(2) :421-427.
  • 7Anyanwu AC, Fischer GW, Plotkina I, et al. Off-pump implant of the Jarvik 2000 ventricular assist device through median sternotomy. Ann ThoracSurg, 2007, 84(4):1405-1407.
  • 8Zhang Y, Xue S, Gui XM, etal. A novel integrated rotor of axial blood flow pump designed with computational fluid dynamics. Artif Organs, 2007, 31(7) :580-585.
  • 9Zhang Y, Zhan Z, Gui XM, et al. Design optimization of an axial blood pump with computational fluid dynamics. ASAIO J, 2008, 54(2):150-155.
  • 10Haft J, Armstrong W, Dyke DB, et al. Hemodynamic and exercise performance with pulsatile and continuous-flow left ventricular assist devices. Circulation, 2007, 116 ( 11 Suppl) :I8-I15.

共引文献60

同被引文献16

引证文献3

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部