期刊文献+

Uncertainty contribution at NIS phototherapy irradiance facility

国家标准研究所光线疗法辐照设备的不确定度分配(英文)
下载PDF
导出
摘要 Phototherapy lamps are the most effective and safest in the treatment of several medical treatments such as bilirubin. They transfer radiant energy expressed by irradiance unit in W/m2 into the skin specifically to achieve a therapeutic reduction in the bilirubin concentration in the blood. National Institute for Standards (NIS) in Egypt builds up a radiometric method to en- sure the competence of phototherapy sources (luminaire) to the standard international and national requirements. Hence, NIS provides traceability to customer through unbroken chain of phototherapy radiometer calibrated as irradiance response in W/m2. Uncertainty model including all parameters accompanied with the measurements is studied.
出处 《Journal of Measurement Science and Instrumentation》 CAS CSCD 2015年第4期332-335,共4页 测试科学与仪器(英文版)
关键词 IRRADIANCE jaundice treatment phototherapy luminaire UNCERTAINTY 辐照 黄疸病治疗 光疗法光源 不确定性
  • 相关文献

参考文献11

  • 1American Academy of Pediatrics. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics, 2004, 114(1): 297-316.
  • 2Facehini F P. Standardizing the calibration of phototherapy devices-a proposal. Jornal de Pediatria, 2001, 77 (2): 67- 74.
  • 3Ferreira A L C, Nascimento R M, Verissimo R C S S. Irradiance of phototherapy equipment in maternity wards in Maeeio. Revista de Latino-Americana de Enferrnagem, 2009, 17(5): 695-700.
  • 4Van den Belt V W, Van der Heide M, Van Imhoff D E, et al. Irradiance levels of phototherapy in jaundiced preterm infants. Archives of Disease in Childhood, 2008: 93.
  • 5Wentworth D P. Neonatal phototherapy-today' s lights,lamps and devices. Infant, 2005, 1(1): 14-19.
  • 6Reda S M, Abd Elbaset Y. Performance evaluation for hyperbilirubinemia phototherapy equipment. Australian Journal of Basic and Applied Sciences, 2013, 7(9): 364-366.
  • 7International Electrotechnical Commission (IEC). Particular requirements for the safety of infant phototherapy equipment. IEC 60601-2-50, 2000.
  • 8Egyptian Organization for Standardization and Quality (EOS). Particular requirements for the safety of the infant Phototherapy equipment. ES 5807. 2007.
  • 9ISO. Guide to the expression of uncertainty in measurement (GUM). International Organization for Standardization, 2008: 98-100.
  • 10United Kingdom Accreditation Service (UKAS). The expression of uncertainty and confidence in measurement. 3rd edition. 2013.