期刊文献+

血小板聚集分析仪LBY-NJ4A与Chrono-log700性能比较 被引量:1

Comparison on the performances of light transmission aggregometry LBY-NJ4A and Chrono-log700
下载PDF
导出
摘要 目的评价并比较全自动血小板聚集分析仪LBY-NJ4A与手工血小板聚集分析仪Chrono-log700的主要性能指标。方法评价并比较LBY-NJ4A和Chrono-log700的精密度、通道一致性、参考范围、抗干扰能力以及携带污染率等主要性能,验证两种仪器的相关性。结果两种仪器的精密度、通道一致性、抗干扰能力、参考范围均符合要求,仪器间的相关性好。另外,LBY-NJ4A携带污染率符合标准,Chrono-log700不存在携带污染率,因为其为一次性加样。结论血小板聚集分析仪LBY-NJ4A与Chrono-log700的性能基本符合临床及实验室需要,但LBY-NJ4A抗干扰能力优于Chrono-log700,且自动化程度较高,人工需求较少,因此更有利于实验室大样本量工作的开展。 Objective To evaluate and compare the main performances of automatic light transmission aggregometry(LBY-NJ4A)and manual light transmission aggregometry(Chrono-log700).Methods The main performances of LBY-NJ4A and Chrono-log700 were evaluated,such as precision,channel consistency,reference range,anti-interference ability and carried pollution rate,and the correlation were verified.Results The precision,channel consistency,anti-interference ability and reference range of the two instruments met the requirements,and two instruments were well correlated.The pollution rate carried by LBY-NJ4A met the standard;Chrono-log700 didn′t have any contamination problem because it only needed disposable tips for loading samples.Conclusion Both LBY-NJ4A and Chrono-log700 are basically accorded with clinical and laboratory requirements.LBY-NJ4A has a better anti-interference quality than Chrono-log700.Additionally,it was highly automated and has less labor requirements,so it might be more favorable for laboratories with a large sample size.
作者 李勤 凌莉琴 刘超男 陈思 周静 LI Qin;LING Liqin;LIU Chaonan;CHEN Si;ZHOU Jing(Department of Laboratory Medicine,West China Hospital,Sichuan University,Chengdu,Sichuan 610041,China)
出处 《国际检验医学杂志》 CAS 2022年第6期641-644,共4页 International Journal of Laboratory Medicine
基金 国家自然科学基金项目(81371878)。
关键词 血小板聚集分析仪 LBY-NJ4A Chrono-log700 性能比较 light transmission aggregometry LBY-NJ4A Chrono-log700 performances comparison
  • 相关文献

参考文献6

二级参考文献31

  • 1Harrison P,FrelingerⅢ AL,Furman MI. Measuring antiplatelet drug effects in the laboratory[J].Thrombosis Research,2007,(03):323-336.
  • 2中国合格评定国家认可委员会.CNAS-CL02医学实验室质量和能力认可准则(ISO15189:2003)[M]北京:中国计量出版社,2006.
  • 3叶应妩;王毓三;申子瑜.全国临床检验操作规程[M]南京:东南大学出版社,200647-81.
  • 4Clinical and Laboratory Standards Institute. EP5A-2 Evaluation of precision performance of quantitative measurent methods-approved guideline[S].USA:Clinical and Laboratory Standards Institute,2004.
  • 5National Committee for Clinical Laboratory Standards. EP-9A2 Method comparison and bias estimation using patoent samples[S].USA:National Committee for Clinical Laboratory Standards,2002.
  • 6郑铁生;鄢盛恺.临床生物化学检验[M]北京:中国医药科技出版社,201075-87.
  • 7姜旭金.临床生物化学检验实验指导[M]北京:中国医药科技出版社,2010121-126.
  • 8National Committee for Clinical Laboratory Standards. EP6-A Evaluation of the linearity of quantitative measurement procedures:a statistical approach,approved guideline[S].USA:NCCLS,2003.
  • 9王学锋;王鸿利.血栓与止血的检测及应用[M]上海:上海世界图书出版公司,200243-69137-140.
  • 10Michelson AD,FrelingerⅢ AL,Furman MI. Current options in platelet function testing[J].American Journal of Cardiology,2006,(10):4-10.

共引文献84

同被引文献17

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部