摘要
目的探讨模拟ECG信号在320排CT冠脉成像中的应用价值。方法收集和分析我院2015-01-01至2021-10-01期间,使用模拟ECG信号成像的20例患者(A组),与同时期,随机抽取的,常规技术成像的20例患者(B组)的冠脉成像结果,实行对照研究。结果A、B两组图像质量主观法评价,图像质量无显著差异(P=0.3758>0.05);A、B两组图像质量客观法评价,升主动脉根部,右冠状动脉近端,左前降支近端,左旋支近端CT值以及升主动脉根部层面噪声均无明显差异(P>0.05);A、B两组辐射剂量对比有显著差异(P<0.01)。结论模拟ECG信号在320排CT冠脉成像中的应用是可行的,值得推广。
Objective To explore the application value of simulated ECG signals in 320-slice CT coronary imaging.Methods Collect and analyze coronary artery imaging of 20 patients(group A)using simulated ECG signal imaging in our hospital from 2015-01-01 to 2021-10-01,At the same time,randomly selected,20 patients(group B)imaging with conventional techniques As a result,a controlled study was carried out.Results The image quality of the two groups A and B were evaluated subjectively,and the images were not significantly different(P=0.3758>0.05);the image quality of the two groups A and B was evaluated by objective methods,ascending aorta root,proximal right coronary artery,and left anterior There were no significant differences in CT values of proximal descending branch,proximal left circumflex branch,and ascending aortic root level noise(P>0.05);there was a significant difference in radiation dose between A and B groups(P<0.01).Conclusion Simulated ECG signal is feasible in 320-slice CT coronary imaging and is worthy of promotion.
作者
成满平
蔡晓庆
牛娟琴
薛巍
陈纲
岳丽娜
杜林芝
Cheng Man-ping;Cai Xiao-qing;Niu Juan-qin;Xue Wei;Chen Gang;Yue Li-na;Du Lin-zhi(Department of Radiodiagnosis,the 940th Hospital of Joint Logistics Support Force,Lanzhou 730050,Gansu Province,China;Department of Cardiology,the 940th Hospital of Joint Logistics Support Force,Lanzhou 730050,Gansu Province,China)
出处
《中国CT和MRI杂志》
2023年第11期77-79,共3页
Chinese Journal of CT and MRI
基金
宁夏自然科学基金项目(2022AAC03716)。