摘要
The accuracy(repeatability and reproducibility) of the iron content analysis of galvanized coating using an X-ray fluorescence spectrometer with an L-spectrum is not better than that of flame atomic absorption spectrometry, sometimes it exceeds the quality control limit.Influences, such as current, voltage, equipment(internal circulating water, 10%CH4+90%Ar, and vacuum) checking, instrument monitoring, sample cleaning, and oper-ators, were investigated by means of 6-sigma and lean operations to improve accuracy.