Evaluation of patient visual comfort and repeatability of refractive values in non-presbyopic healthy eyes

Evaluation of patient visual comfort and repeatability of refractive values in non-presbyopic healthy eyes
下载PDF
导出
摘要 AIMTo evaluate the intra-operator repeatability in healthy subjects using the WAM-5500 auto-kerato/refractometer and the iTrace aberrometer, to compare the refractive values and the subjective refraction obtained with both devices and to determine which of these three spherocylindrical corrections allows the subject to achieve the best visual comfort.METHODSForty-two non-presbyopic healthy eyes of 42 subjects were enrolled in this prospective study. Refractive values were compared, evaluating the repeatability, the relationship between the methods and the best visual comfort obtained.RESULTSSphere, cylinder and axis results showed good intraclass correlation coefficients (ICC); the highest ICC was obtained using the spherical refraction with the autorefractometer and the aberrometer, achieving levels of 0.999 and 0.998, respectively. The power vector (PV) was calculated for each refraction method, and the results indicated that there were no statistically significant differences between them (P>0.05). Direct comparison of PV measurements using the three methods showed that aberrometer refraction gave the highest values, followed by the subjective values; the autorefractometer gave the lowest values. The subjective method correction was most frequently chosen as the first selection. Equal values were found for the autorefractometer and the aberrometer as the second selection.CONCLUSIONThe iTrace aberrometer and the WAM-5500 auto-kerato/refractometer showed high levels of repeatability in healthy eyes. Refractive corrections with the aberrometer, the autorefractometer and subjective methods presented similar results, but spherocylindrical subjective correction was the most frequently selected option. These technologies can be used as complements in refractive evaluation, but they should not replace subjective refraction. AIMTo evaluate the intra-operator repeatability in healthy subjects using the WAM-5500 auto-kerato/refractometer and the iTrace aberrometer, to compare the refractive values and the subjective refraction obtained with both devices and to determine which of these three spherocylindrical corrections allows the subject to achieve the best visual comfort.METHODSForty-two non-presbyopic healthy eyes of 42 subjects were enrolled in this prospective study. Refractive values were compared, evaluating the repeatability, the relationship between the methods and the best visual comfort obtained.RESULTSSphere, cylinder and axis results showed good intraclass correlation coefficients (ICC); the highest ICC was obtained using the spherical refraction with the autorefractometer and the aberrometer, achieving levels of 0.999 and 0.998, respectively. The power vector (PV) was calculated for each refraction method, and the results indicated that there were no statistically significant differences between them (P>0.05). Direct comparison of PV measurements using the three methods showed that aberrometer refraction gave the highest values, followed by the subjective values; the autorefractometer gave the lowest values. The subjective method correction was most frequently chosen as the first selection. Equal values were found for the autorefractometer and the aberrometer as the second selection.CONCLUSIONThe iTrace aberrometer and the WAM-5500 auto-kerato/refractometer showed high levels of repeatability in healthy eyes. Refractive corrections with the aberrometer, the autorefractometer and subjective methods presented similar results, but spherocylindrical subjective correction was the most frequently selected option. These technologies can be used as complements in refractive evaluation, but they should not replace subjective refraction.
出处 《International Journal of Ophthalmology(English edition)》 SCIE CAS 2015年第5期1031-1036,共6页 国际眼科杂志(英文版)
基金 Supported by the Government of Aragon,Group B99 Fund of Health Research from Institute of Health Carlos Ⅲ(Spanish Ministry of Health)PS0901854 and PI13/01124 University of Zaragoza grant FPUZ-2011-BIO-02
关键词 REPEATABILITY optical correction aberrometer autorefractometer subjective refraction repeatability optical correction aberrometer autorefractometer subjective refraction
  • 相关文献

参考文献1

二级参考文献19

  • 1Smith G. The accommodative resting states, instrmnent accommodation and their measurement. J Mod Optic, 1983, 30: 347 -359.
  • 2Rosenfield M, Ciuffreda KJ. Effect of surround propinquity on the open-loop accommodative response. Invest Ophthalmol Vis Sci, 1991, 32: 142-147.
  • 3Barry JC,Konig HH. Non-cycloplegic screening for amblyopia via refractive findings with the Nikon Retinomax hand held autorefractor in 3 year old kindergarten children. Br J Ophthalmol, 2001, 85 : 1179-1182.
  • 4Mallen EA,Wolffsohn JS, Gilmartin B, et al. Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in adults. Ophthalmic Physiol Opt, 2001, 21 : 101-107.
  • 5Davies LN, Mallen EA,Wolffsohn JS, et al. Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor. Optom Vis Sci, 2003, 80 : 320-324.
  • 6Sheppard AL, Davies LN. Cfinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500. Ophthalmic Physiol Opt, 2010, 30: 143-151.
  • 7Gwiazda J,Weber C. Comparison of spherical equivalent refraction and astigmatism measured with three different models of autorefractors. Optom V-is Sci, 2004, 81 : 56-61.
  • 8Cordonnier M, Dramaix M. Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism. Br J Ophthalmol, 1999, 83: 157-161.
  • 9Wesner MF, Miller RJ. Instrument myopia conceptions, misconceptions, and influencing factors. Doc Ophthalmol, 1986, 62: 281-308.
  • 10Miwa T. Instrument myopia and the resting state of accommodation. Optom Vis Sci, 1992, 69: 55-59.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部