CT机房楼板动力特性分析与减振改造设计 被引量:3

Dynamic analysis and vibration reduction transformation design of CT room floor
原文传递
导出
摘要 针对医院CT机房CT设备运行时引起的振动问题,通过对其工作区域动力特性的测试,分析振动环境对工作舒适度及楼板结构安全的影响,并在此基础上对CT机房楼板进行减振改造设计。采取在楼板下安装可调节刚度调谐质量阻尼器(TMD)以及粘钢加固措施,既保证了楼板的承载能力,又有效地减小了设备运行引起的振动。可为类似工程振动分析及控制设计提供参考和依据。 Aimed at the vibration caused by CT equipments in CT room of a hospital,through testing the dynamic response in CT equipments working region,the impact of vibration circumstance on working comfort level and the floor safety was analyzed.The vibration reduction transformation design was made on this basis.By installing adjustable stiffness tuned mass dampers(TMD) and using steel-bonded reinforcement measure,it could guarantee the floor' s bearing capacity and reduce effectively the vibration caused by running equipment.Reference can be offered for similar project vibration analysis and manage design.
出处 《建筑结构》 CSCD 北大核心 2012年第S1期605-608,共4页 Building Structure
关键词 CT设备 动力特性 减振改造 TMD CT equipments dynamic response vibration reduction transformation TMD
  • 相关文献

参考文献7

二级参考文献18

  • 1孙利民,闫兴非.人行桥人行激励振动及设计方法[J].同济大学学报(自然科学版),2004,32(8):996-999. 被引量:124
  • 2房良,蓝宗建,周伟.TMD系统参数有效域的研究[J].工业建筑,2004,34(10):10-12. 被引量:6
  • 3聂建国,孙彤,温凌燕,刘静.某会展中心大跨交叉钢-混凝土组合梁系楼盖设计[J].建筑结构学报,2004,25(6):123-125. 被引量:23
  • 4Zheng X, Brownjohn J M W. Modeling and simulation of human-floor system under vertical vibration [ C ] //Proceedings of SPIE, Smart Structures and Materials. Newport Beach : International Society for Optical Engineering, 2001: 513-520.
  • 5Setareh M, Hanson R D. Tuned mass dampers to control floor vibration from humans [ J ]. Journal of Structural Engineering, 1992, 118(3):741-762.
  • 6Hurry W C. Dynamic analysis of structural systems using component modes [ J ]. AIAA Journal, 1965, 3 (4) : 678-685.
  • 7Young P. Improved floor vibration prediction methodologies [ C ]// ARUP Vibration Seminar. London :ARUP, 2001.
  • 8Murray T M, Allen D E, Ungar E E. Floor vibrations due to human activity, AISC design guide 11 [ S]. Chicago: American Institute of Steel Construction, 1997.
  • 9Pavic A, Reynolds P, Waldron P, et al. Dynamic modeling of post-tensioned concrete floors using finite element analysis [ J]. Finite Elements in Analysis & Design, 2001, 37(4): 305-323.
  • 10E1-Dardiry E, Wahyuni E, models of a long-span fiat Ji T, et al. Improving FE concrete floor using natural frequency measurements [ J ]. Computers and Structures, 2002, 80 ( 27-30 ) : 2145-2156.

共引文献182

同被引文献16

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部