Synergy CBCT两种配准方式对脑肿瘤放疗摆位误差的研究 被引量:2

A Study on the Set-Up Errors of Brain Tumors by Two Kinds of Registration Methods of Synergy CBCT
下载PDF
导出
摘要 目的:分析脑肿瘤的影像引导放射治疗中,Synergy加速器里的CBCT两种影像配准方式的摆位误差。方法:选择100例脑肿瘤患者,在治疗前进行CBCT扫描,将得到的扫描影像与计划CT影像对比,分别采用骨配准与灰度配准方式。记录这两种配准方式在X、Y、Z三个线性方向上的误差值,并采用配对t检验对数据进行两两比较,计算差别是否达到显著性水平。结果:采用骨配准在X、Y、Z方向上的误差值分别是(-0.051±0.136)cm、(-0.059±0.130)cm、(0.057±0.132)cm,灰度配准在X、Y、Z方向上的误差值分别是(-0.055±0.135)cm、(-0.065±0.129)cm、(0.065±0.129)cm。三个方向上统计学分析表明这两种配准方式的数据差异达到显著性水平。结论:初步研究显示在脑肿瘤的影像引导放疗中,CBCT的骨配准精度比灰度配准精度更高。 Purpose:To analyze the set-up errors of two kinds of image registration methods in CBCT of Synergy accelerator in the image guided radiotherapy of brain tumors.Methods:A total of 100 patients with brain tumor were enrolled for CBCT scan before first treatment,compared the obtained image with planned CT image,bone and grayscale registration were used respectively.The error values of two kinds of registration methods were recorded in X,Y and Z axes.Paired t-test was used to evaluate the differences.Results:The error values of bone registration in X,Y and Z axes were(-0.051±0.136)cm,(-0.059±0.130)cm and(0.057±0.132)cm respectively,while(-0.055±0.135)cm,(-0.065±0.129)cm and(0.065±0.129)cm with gray-scale registration.The difference between the two methods in three directions was with statistical significance.Conclusion:The accuracy of bone registration is higher than that of gray-scale registration in CBCT of brain tumors.
作者 朱俐颖 胡小洋 孙磊 盛晓芳 汪洋 ZHU Liying;HU Xiaoyang;SUN Lei;SHENG Xiaofang;WANG Yang(Department of Radiation Oncology,Huashan Hospital,Fudan University,Shanghai 201399,China;Department of Radiation Oncology,Shanghai Gamma Hospital;Cyberknife Center,Huashan Hospital,Fudan University)
出处 《中国医学计算机成像杂志》 CSCD 北大核心 2022年第3期317-321,共5页 Chinese Computed Medical Imaging
关键词 放疗 影像引导放疗 图像配准 脑肿瘤 Radiotherapy Image-guide radiotherapy Image registration Brain neoplasms
  • 相关文献

参考文献9

二级参考文献93

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2.调强适形放射治疗[A].胡逸民.肿瘤放射物理学[C].北京:原子能出版社,1999.538-612.
  • 3Hanley J,Debois MM,Mah D,et al.Deep inspiration breath-hold technique for lung tumors:the potential value of target immobilization and reduced lung density in dose escalation.Int J Radiat Oncol Biol Phys,1999,45:603-611.
  • 4Ramsey CR,Scaperoth D,Arwcod D,et al.Clinical efficacy of respiratory gated conformal radiation therapy.Med Dosim,1999,24:115-119.
  • 5Shirato H,Shimizu S,Kitamura K,et al.Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.Int J Radiat Oncol Biol Phys,2000,48:435-442.
  • 6Keall PJ,Chen GTY,Joshi S,et al.Time-the fourth dimension in radiotherapy (ASTRO panel discussion).Int J Radiat Oncol Biol Phys,2003,57:S8-S9.
  • 7Yu CX,Jaffray DA,Wong JW.The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation.Phys Med Biol,1998,43:91-104.
  • 8de Neve W,Heuvel F,De Beukeleer M,et al.Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch.Radiother Oncol,1992,24:45-54.
  • 9Hu YM,DM JR,Hu B.The precision and accuracy analysis of target position in X-ray stereotactic radiotherapy for extracranial lesions.In:Internation symposium of stereotactic radiosurgery (SRS/SRT).Taejon:Chungnam National University Hospital,1997.
  • 10Yan D,Vicini F,Wong J,et al.Adaptive radiation therapy.Phys Med Biol,1997,42:4123-4132.

共引文献252

同被引文献25

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部