对水中总氮测定的前处理消解方法进行了改进,用烘箱法消解代替高温蒸汽消解;对两种前处理方法分析结果的线性关系、准确度、精密度、加标回收率、检出限等进行了验证和对比。结果表明,烘箱法前处理的分析结果满足测试要求。
The pretreatment digestion method for the determination of total nitrogen in water is renovated,and the oven digestion method is instead of the high temperature steam digestion method.The linear relationship,accuracy,precision,spike recovery rate,detection limit,etc.of the analysis results by two pretreatment methods are verified and compared.The results show that the oven digestion method can meet the requirements in quality.
作者
孙悦嘉
SUN Yue-jia(Suzhou Kunhuan Detection Technology Co.,Ltd.,Kunshan 215300,China)
出处
《现代化工》
CAS
CSCD
北大核心
2022年第S02期424-426,共3页
Modern Chemical Industry
关键词
总氮
前处理
烘箱法
准确度
精密度
检出限
total nitrogen
pretreatment
accuracy
precision
detection limit