高能X射线下水等效材料对水的等效性研究 被引量:2

Study on the Equivalence of Water Equivalent Materials to Water under High-energy X-ray
下载PDF
导出
摘要 针对水等效材料校准医用加速器中存在的问题,通过对比法测量水体模和3种水等效材料校准医用加速器的剂量学差异,分别给出PTW 30013和IBA FC65-G两种类型探测器在校准深度为5 cm和10 cm的修正因子。针对加速器日稳定性校准的必要性,通过近7个月的日稳定性测量结果,给出加速器漂移量和引起漂移的可能原因;针对水等效材料在相对测量中对水的等效性,通过测量12个辐射野的输出因子,得出水等效材料在相对测量中与水一致。最后给出水等效材料在高能X射线下水吸收剂量校准时的建议和注意事项。 The problems in the calibration of accelerator using water equivalent materials were described. Three commonly used water equivalent materials were compared to water phantom in measurements of absorbed dose to water. Two Farmer-type cylindrical chamber,PTW 30013 and IBA FC65-G,were used in measurements. The calibration factors of absolute dose between water phantom and water equivalent materials at two depth,5 cm and 10 cm,were listed. Through the results of daily stability nearly seven months,the necessity of daily calibration of the accelerator was explained. The dose drift of accelerator and the possible causes were given. By measuring the output factor of twelve radiation fields,the water equivalent material is consistent with water in the relative measurement. Finally,the suggestions and precautions for the water equivalent material in the calibration of high energy X-ray were given.
作者 王志鹏 邢书明 王坤 金孙均 范富有 刘福斌 WANG Zhi-peng;XING Shu-ming;WANG Kun;JIN Sun-jun;FAN Fu-you;LIU Fu-bin(School of Mechanical Electronic and Control Engineering,Beijing Jiaotong University,Beijing 100044,China;National Institute of Metrology,Beijing 100029,China)
出处 《计量学报》 CSCD 北大核心 2020年第2期257-262,共6页 Acta Metrologica Sinica
基金 国家重点研发计划(2018YFF0212403,2016YFF0200804).
关键词 计量学 水等效材料 日稳定性 高能X射线 绝对剂量 输出因子 metrology water equivalent material daily stability high energy X-ray absolute dose output factor
  • 相关文献

参考文献4

二级参考文献31

  • 1IAEA. Absorbed dose determination in external beam radiotherapy. IAEA Report TRS-398_ IAEA. Vienna: IAEA, 2001.
  • 2Gerbi B, Antolak JA, Deibel FC, et al. Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25. Med Phys, 2009,36: 3239- 3279.
  • 3Tello VM, Tailor RC, Hanson WF. How water equivalent are water-equivalent solid materials for output calibration of photon and electron beams ? Med Phys, 1995,22 : 1177-1189.
  • 4Liu L, Prasad SC, Bassand DA. Evaluation of two water- equivalent phantom materials for output calibration of photon and electron beams. Med Dosi,2003,4:267-269.
  • 5张红志.电离辐射吸收剂量的测量//胡逸民,主编.肿瘤放射物理学.北京:原子能出版社,1999:64-67.
  • 6Ding DX, Rogers DWO, Mackie TR. Mean energy, energy range relationships and depth-scaling factors for clinical electron beams. Med Phys, 1996,23:361-376.
  • 7Pruitt JS, Loevinger R. The photon-fluence scaling theorem for Compton-scattered radiation. Med Phys,1982 ,9 :176-179.
  • 8Seuntjens J, Olivares M, Evans M, et al. Absorbed dose to water reference dosimetry using solid phantoms in the ontext of absorbed- dose protocols. Med Phys,2005,32:2945-2953.
  • 9杨瑞峰,黄辉,何涛.Siemens primus E医用直线加速器的剂量校准[J].医疗卫生装备,2009,30(4):115-117. 被引量:6
  • 10徐宜武,林益匡.不规则射野的百分深度剂量计算[J].中国医疗设备,2011,26(2):39-41. 被引量:1

共引文献16

同被引文献15

引证文献2

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部