福州地区乳腺数字X射线摄影辐射剂量调查 被引量:2

Investigation on radiation dose of digital mammography in Fuzhou
原文传递
导出
摘要 目的评估福州地区乳腺数字X射线摄影的辐射剂量水平,为制定或修订福建省或全国医疗照射指导水平提供基础数据。方法选取福州地区的4家医院160例女性受检者,对头尾位(CC位)投照和侧斜位(MLO位)投照的临床曝光参数进行调查分析。结果本次调查的乳房压迫厚度中位数为44 mm;乳腺平均剂量的中位数为1.75 mGy;MLO位检查时的管电压、管电流时间积、入射体表空气比释动能、乳房压迫厚度、乳腺平均剂量均高于CC位,但两者比较差异无统计学意义(P>0.05);随着乳房压迫厚度的增加,乳腺平均剂量也呈增加趋势,且差异具有统计学意义(P<0.05)。结论福州地区MLO位和CC位的乳房压迫厚度均略高于全国的调查水平;乳腺平均剂量75%位值为2.25 mGy,高于国家标准,今后将进一步加强对乳腺平均剂量影响相关因素的研究。 Objective To evaluate the radiation dose level of digital mammography in Fuzhou.Methods A total of 160 female patients were selected from 4 hospitals in Fuzhou,the exposure parameters of craniocaudal(CC)and mediolateral oblique(MLO)were investigated.Results The median breast compression thickness was 44 mm,the median breast dose was 1.75 mGy.Tube voltage,tube current time product,air specific release kinetic energy,compression thickness of breast and average dose of breast in MLO were all higher than those in CC,but there was no statistically significant difference between them(P>0.05).With the increase of breast compression thickness,the average breast dose also increased,and the difference was statistically significant(P<0.05).Conclusion The breast compression thickness at MLO and CC in Fuzhou was slightly higher than the national survey level;the median breast dose was 1.75 mGy,lower than the national standard,and the mean breast dose level was positively correlated with the thickness of breast compression.Further research on the related influencing factors of average breast dose should be strengthened in the future.
作者 王晓娜 苏丹华 王心韬 施文华 刘祥铨 李文最 WANG Xiaona;SU Danhua;WANG Xingtao;SHI Wenhua;LIU Xiangquan;LI Wenzui(The Center for Disease Control and Prevention of Fuzhou,Fuzhou,Fujian,350004,China)
出处 《河南预防医学杂志》 2021年第3期228-230,共3页 Henan Journal of Preventive Medicine
关键词 乳腺数字X射线摄影 辐射剂量 调查 Digital mammography Radiation dose Investigation
  • 相关文献

参考文献6

二级参考文献71

  • 1郑钧正,朱志贤,唐文祥,韩发明,何韦川.乳腺X射线摄影所致受检者乳腺平均剂量调查研究[J].中华放射医学与防护杂志,2001,21(S1). 被引量:9
  • 2卞丽琴,梅红,宋少娟.全数字化乳腺摄影技术的临床应用[J].医学影像学杂志,2006,16(6):653-654. 被引量:7
  • 3岳保荣,尉可道,刘澜涛,王建超.正确选用乳腺普查放射技术[J].中华放射医学与防护杂志,2006,26(6):547-548. 被引量:11
  • 4Law J, Faulkner K. Radiation benefit and risk at the assessment stage of the UK Breat Screening Programme. BJR, 2006,79: 479-482.
  • 5Heyes GJ. Enhanced biological effectiveness of low energy X-ray and implications for the UK breast screening programme. BJR, 2006 79: 195-200.
  • 6Suryanrayanan S, Karellas R, Vedantham S, et al. Hat-panel digital mammography system: contrast-detail comparison between screen-film radiographs and hard-copy images, Radiology, 2002,225: 801-808.
  • 7Robert M, Gagne BD, Myers KJ. Toward objective and quantitative evaluation of imaging systems using image of phantom. Med Phys, 2006, 33( 1 ) : 83-95.
  • 8Thomas JA, Chakrabarti K, Romanyukha A, et al. Contrast-detail phantom scoring methodology, Med Phys, 2005,32(3):808-14.
  • 9Buhr E, Gunther-Kohfahl S, Neitzel U. Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image. Med Phys, 2003,30 ( 9 ) : 2323-2331.
  • 10Samei E, Ranger NT, Dobbins JT, et al. Intercomparison of methods for image quality characterization.Ⅰ. Modulation transfer function. Med Phys, 2006,33(5) : 1454-1465.

共引文献38

同被引文献17

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部