期刊文献+

污泥浓度(MLSS)在线分析仪示值误差的不确定度评定

Uncertainty Evaluation of Indication Error of MLSS Online Analyzer
原文传递
导出
摘要 以HACH生产的sc200污泥浓度(MLSS)在线分析仪为例,对影响其测量结果不确定度的各个分量进行分析,由于目前没有对应的检定规程和校准规范,也没有适用的国家标准物质,因此,采用过滤称重法对仪器的示值误差进行不确定度评定。以期为企业污水的排放监测和工艺控制提供计量数据支撑。 This paper takes the SC200 analyzer on-Line of sludge concentration(MLSS)of HACH as an example,All of affecting the measurement result of each component is analyzed.because there is no corresponding verification regulations and calibration specifications at present,also does not have the applicable national standard substances,therefore,In order to provide measurement data support for wastewater discharge monitoring and process control,the article adopts filter weighing method for uncertainty of error of the instrument.
作者 马振奇 朱之枫 宋少磊 Ma Zhenqi;Zhu Zhifeng;Song Shaolei(Henan Institute of Metrology,Zhengzhou 450003,China;Luohe Technology Extension Center of Agricultural Water,Luohe 462002,China)
出处 《环境工程》 CAS CSCD 北大核心 2023年第S02期955-956,共2页 Environmental Engineering
关键词 污泥浓度 示值误差 不确定度 sludge concentration indicating value error uncertainty
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部