期刊文献+

低管电压结合迭代重建在下肢动脉CT血管成像中的临床应用 被引量:6

Clinical Application of Low Tube Voltage Combined with Iterative Reconstruction in Lower Extremity Arterial CTA
下载PDF
导出
摘要 目的:探讨80 kV管电压结合迭代重建技术在下肢动脉CT血管成像中的应用价值。方法:将40例拟行下肢CTA检查的患者随机分为两组,实验组扫描及注射参数为80 kVp,对照组为100 k Vp。对比两组辐射剂量、对比剂剂量和图像质量。结果:对照组的有效辐射剂量和对比剂剂量(95 mL)明显高于实验组,差异具有统计学意义(P<0.001);两组血管强化CT值和图像质量主观评分均无统计学差异(P均大于0.05)。结论:80 kV管电压结合迭代重建技术的CT扫描技术应用于下肢动脉CTA成像,在保证图像质量的前提下能显著降低辐射剂量和对比剂剂量,具有重要应用推广价值。 Objective:To explore the application value of 80 kV tube voltage combined with iterative reconstruction technology in CT angiography of lower extremity artery.Methods:40 patients were randomly divided into two groups.The parameters of scanning and injection were 80 kV in the experimental group and 100 kV in the control group.The radiation dose,contrast dose and image quality were compared between the two groups.Results:The effective radiation dose and contrast dose(95 mL)of the control group were significantly higher than those of the experimental group,the difference was statistically significant(P<0.001);there was no statistical difference between the two groups in the CT value of vascular enhancement and the subjective score of image quality(P>0.05).Conclusion:The CT scanning technology of 80 kV tube voltage combined with iterative reconstruction technology can significantly reduce the radiation dose and contrast agent dose on the premise of ensuring the image quality,which has important application and promotion value.
作者 张俊 刘佳怿 梁志鹏 ZHANG Jun;LIU Jiayi;LIANG Zhipeng(Department of Radiology,Sir RunRun Hospital affiliated of Nanjing Medical University,Nanjing 211100,China)
出处 《CT理论与应用研究(中英文)》 2020年第1期55-62,共8页 Computerized Tomography Theory and Applications
关键词 下肢动脉 CT血管成像 低管电压 迭代重建 lower extremity artery CT angiography low tube voltage iterative reconstruction
  • 相关文献

参考文献3

二级参考文献21

  • 1D.Utsunomiya,S.Oda,Y.Funama,K.Awai,T.Nakaura,Y.Yanaga,程悦.比较外周动脉疾病病人的标准与低管电压MDCT血管成像[J].国际医学放射学杂志,2011(1):94-95. 被引量:8
  • 2MC Lange, Zetola VF, Teive HA, et al. Cerebrotendinous xanthomatosis: report of two Brazilian brothers [ J ]. Arq Neuropsiquiatr, 2004,62 : 1085-1089.
  • 3M He, Qin X, Cui Y, et al. Prevalence of unrecognizedlower extremity peripheral arterial disease and the associated factors in chinese hypertensive adults[J]. Am J Cardiol, 2012,110 : 1692-1698.
  • 4Young Y, Chi KK, Ajaka J, et al. Complications with out-patient angiography and intelventional procedures [ J ]. Ca rdiovasc Intervent Radiol, 2002,25 : 123-126.
  • 5Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure [ J]. N Engl J Med, 2007, 357 : 2277- 2284.
  • 6Davies HE, Wathen CG, Gleeson FV. The risks of radia- tion exposure related to diagnostic imaging and how to mini- raise them [ J ]. BMJ, 2011,342 : d947.
  • 7Qi L, Zhao Y, Zhou CS, et al. Image quality and radiation dose of lower extremity CT angiogTaphy at 70 kVp on an in- tegrated circuit detector dual-source computed tomography [J].Aeta Radiol, 2015,56:659-665.
  • 8郭濴,葛英辉,史大鹏,郭英,宁培钢,温泽迎,张守华.64层螺旋CT自动毫安技术对降低下肢动脉血管成像辐射剂量的研究[J].中华放射医学与防护杂志,2009,29(6):624-626. 被引量:15
  • 9涂永波,兰永树,舒健,程广金,唐烨真,唐光才.螺距对下肢CTA辐射剂量与图像质量影响的研究[J].中国医学影像学杂志,2011,19(4):300-303. 被引量:20
  • 10代敏,郑文龙,吴爱琴,何瑶瑶,胡将珍,严志汉.下肢CT血管成像低kVp和低mAs扫描对比研究[J].影像诊断与介入放射学,2013,22(2):145-148. 被引量:5