期刊文献+

Parameters Optimal Selection during Pediatric 18F-FDG PETCT Examination

原文传递
导出
摘要 Objective:Up to now,pediatric 18F-FDG dose and acquisition durations are generally based on coarse extrapolation from adult guidelines.This study sought to determine whether shorter acquisition durations or a lower 18F-FDG injected activity could be used during pediatric 18F-FDG PET/CT examinations while maintaining diagnostic utility.Methods:36 whole-body 18F-FDG PET/CT examinations were performed on 36 patients(weight,13-89 kg;46.51±5.63 kg,age range,3-14 y,9.22±3.16)with a weight-based injected activity(5.3 MBq/kg[0.144 mCi/kg]),fixed acquisition durations 180 S/FOV,Vip record acquisition mode using Discovery STE.For each examination,the Vip-mode data were truncated to form multiple datasets with shorter acquisition durations down to a minimum of 60 S/FOV(i.e.,60,80,100,120,140,160 S/FOV data were formed from single 180 S/FOV acquisition).168 image volumes were generated,randomized,and reviewed in a masked manner with corresponding CT image volumes by 6 radiologists.Overall,subjective adequacy and objective lesion detection accuracy by body region were evaluated.Results:All examinations with maximum acquisition duration were graded as adequate and were used as the reference standard for detection accuracy.For patients more than 30 kg,when acquisition duration was more than 120 S/FOV,all PET/CT examinations was graded as adequate for clinical tasks,whereas,acquisition duration was reduced less than 120 S/FOV,lesion detection became less accurate.For patients less than 30 kg,Lesion detection accuracy was perfect for acquisition times between 140 S/FOV and 180 S/FOV for all regions of the body.However,lesion detection became less accurate when imaging acquisition time was reduced less than 140 S/FOV.Conclusion:When GE Discovery STE PET/CT was applied during Pediatric PET/CT examination,Using decreased acquisition times as a surrogate for 18F-FDG dose,18F-FDG dose can be reduced by approximately 33.33%,when patients lager than 30 kg were scanned for 180 S/FOV,for patients less than 30 kg,18F-FDG dose can be reduced by approximately 22.22%without a loss of diagnostic utility.Reduction of overall scan time potentially reduces motion artifacts,improves patient comfort,and decreases length of sedation.Alternatively,decreased 18F-FDG dose minimizes radiation risk.
出处 《Chinese Journal of Biomedical Engineering(English Edition)》 2019年第4期139-144,共6页 中国生物医学工程学报(英文版)
基金 the Fund Program for the Scientific Activities of Selected Returned Overseas Professionals in Shanxi Province for Support
  • 相关文献

参考文献1

二级参考文献15

  • 1范西江,金榕兵,张荷,李奇明,唐维佳.回旋加速器工作参数的选择[J].医疗设备信息,2004,19(9):46-47. 被引量:6
  • 2Zuhayra M, Alfteimi A, Forstner CV, et al. New approach for the synthesis of [ 18F]fluoroethyltyrosine for cancer imaging: simple, fast, and high yielding automated synthesis[J]. Bioorg Med Chem, 2009, 17 (21): 7441-7448.
  • 3Yu S. Review of F-FDG Synthesis and Quality Control[J]. Biomed Imaging Interv J, 2006, 2(4): e57.
  • 4Eberl S, Eriksson T, Svedberg O, et al. High beam current operation of a PETtrace cyclotron for 18F- production[J]. Appl Radiat Isot, 2012, 70(6): 922-930.
  • 5Waiters LR, Martin KJ, Jacobson MS, et al. Stability evaluation of (18)F-FDG at high radioactive concentrations[J]. J Nucl Med Technol, 2012, 40(1): 52-56.
  • 6Calandrino R, del Vecchio A, Parisi R, et al. Measurements and evaluation of the risks due to external radiation exposures and to intake of activated elements for operational staff engaged in the main- tenance of medical cyclotrons[J]. Radiat Prot Dosimetry, 2010, 139(4): 477-482.
  • 7McCarthy M, Yuan JB, Campbell A, et al. 18F-fluorodeoxyglucose positron emission tomography imaging in brain tumours: the Western Australia positron emission tomography/cyclotron service experience [J]. J Med Imaging Radiat Oncol, 2008, 52(6): 564-569.
  • 8Bowden L, Vintro LL, Mitchell PI, et al. Radionuclide impurities in proton-irradiated [180]H20 for the production of 18F: activities and distribution in the [18F]FDG synthesis process[J]. Appl Radiat Isot, 2009, 67(2): 248-255.
  • 9Zanzonico P, Dauer L, St Germain J. Operational radiation safety for PET-CT, SPECT-CT, and cyclotron facilities[J]. Health Phys, 2008, 95 (5): 554-570.
  • 10Pant GS, Senthamizhchelvan S. Initial experience with an 11 MeV self-shielded medical cyclotron on operation and radiation safety[J]. J Med Phys, 2007, 32(3): 118-123.

共引文献5