摘要
目的:对比观察过氧乙酸与甲醛2种消毒剂在透析器复用中的效果。方法:透析器复用采用0.3%过氧乙酸消毒的为A组,采用2%甲醛消毒的为B组,分别比较2组细菌培养、病人透析后血尿素氮、肌酐值及透析器血室容积(FBV)等指标。结果:2组上述3项指标均无统计学差异(P均>0.05)。结论:过氧乙酸可以取代甲醛作为透析器复用消毒剂,且安全、有效。
Objective To observe the effect of peracepic acid and formaldehyde used to sterilize dialyzer. Method Group A used 0.3% peracepic acid to sterilize dialyzer and group B used 2% formaldehyde to sterih'ze it. Then compared indexes such as germ cultivation, post-dialyze BUN, Cr, FBV. Result There was no significant difference between two groups in three indexes mentioned above (P < 0.05). Conclusion Peracepic acid can replace formaldehyde to sterilize dialyzer and it is safe as well as effective.
关键词
过氧乙酸
甲醛
透析器
消毒
Peracepic acid
Formaldehyde
Dialyzer
Sterilization