摘要
本文介绍了BSL-3实验室中实现负压控制的送、排风系统以及送、排风量的具体计算,同时还详细阐述了实现负压控制的具体方案,为BSL-3实验室负压系统的设计提供了理论依据;另外还从节能的角度介绍了BSL-3实验室中使用全热交换器来实现能源消耗最优化。
The article presents the system of air blow and ventilation of negative pressure control andthe calculation of air quantity in Class III Bio-safety lab, furthermore it introduce the concrete schemeof negative control, which offers the theory evidence for design of the system of negative pressurecontrol in Class III Bio-safety lab. In addition the article introduces volatile blow control and heatexchanger using in the system of negative pressure control in Class III Bio-safety lab on the point ofenergy conservation.
出处
《中国医院建筑与装备》
2005年第6期18-23,共6页
Chinese Hospital Architecture & Equipment
关键词
BSL-3实验室
负压控制
送风
排风
系统
控制方案
Class Ⅲ Bio-safety lab negative pressure control air blow ventilation system thescheme of control