双腔起搏器自身房室传导搜索功能临床应用 被引量:3

Clinical Observation of Dual Chamber Pacemaker With Autointrinsic Conduction Search The authors’units
下载PDF
导出
摘要 目的评价具有自身房室传导搜索功能(Autointrinsic Conduction Search,AICS)双腔起搏器减少右心室起搏比例的效果。方法选择18例病窦综合征患者(排除Ⅲ度房室传导阻滞患者),观察应用AICS功能前后患者自身房室传导的比例变化。结果除2例失访外,16例在自身房室传导功能打开后,总心室起搏比例明显下降(5.43±1.7%vs29.5±10.05%,P<0.05),心室起搏百分比>10%的患者比例亦明显减少(5/16例vs10/16例)。结论自身房室传导自动搜索功能可以有效降低右心室起搏的比例,因而可以减少起搏器的能量消耗,改善患者血流动力学效应。 Objective To assess the efficacy of the decrease of right ventricular pacing percentage by the pacemaker with Autointrinsic Conduction Search( AICS ). Methods 18 patients implanted dual chamber pacemakers with AICS functions were followed up(patients with Ⅲ0AVB were excluded)and the percentage of right ventricular pacing were interrogated as the AICS functions were programmed "on" or "off". Results Excepted 2 patients withdrew from the study, the percentage of 16 patients’ right ventricular pacing decreased...
出处 《中国医疗前沿(学术版)》 2008年第4期15-17,共3页 China Healthcare Innovation
关键词 双腔起搏器 自身房室传导搜索 随访 Dual chamber pacemaker Autointrinsic Conduction Search( AICS ) Follow-up
  • 相关文献

参考文献2

共引文献17

同被引文献39

  • 1Connolly SJ, Kerr CR, Gent M, et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes [ J ]. N Engl J Med,2000,342 ( 19 ) : 1385 -1391.
  • 2Lamas GA, Lee KL,Sweeney MO, et al. Ventricular pacing or dualchamber pacing for sinus-node dysfunction [ J ]. N Engl J Med, 2002,346 ( 24 ) : 1854 - 1862.
  • 3Toff WD,Camm AJ,Skehan JD,et al. Single-chamber versus dualchamber pacing for high-grade atrioventricular block [ J ]. N Engl J Med,2005,353 ( 2 ) : 145 - 155.
  • 4Albertsen AE,Nielsen JC, Poulsen SH, et al. Biventrieular pacing preserves left ventricular performance in patients with high-grade atrio-ventricular block: a randomized comparison with DDD ( R ) pacing in 50 consecutive patients [J]. Europace, 2008,10 ( 3 ) : 314 - 320.
  • 5Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction [ J ]. Circulation, 2003,107 (23) :2932 -2937.
  • 6Ptirerfellner H, Brandt J, Israel C, et al. Comparison of two strategies to reduce ventricular pacing in pacemaker patients [J]. Pacing Clin Electrophysiol,2008,31 (2) : 167 - 176.
  • 7Sweeney MO, Bank AJ, Nsah E, et al. Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease [ J ]. N Engl J Med ,2007,357 ( 10 ) : 1000 - 1008.
  • 8Gillis AM, Purerfellner H, Israel CW, et al. Reducing unnecessary right ventricular pacing with the managed ventricular pacing mode in patients with sinus node disease and AV block[J]. Pacing Clin Electrophysiol,2006,29 ( 7 ) :697 - 705.
  • 9Milasinovie G, Tscheliessnigg K, Boehmer A, et al. Percent ventricular pacing with managed ventricular pacing mode in standard pacemaker population [ J ]. Europace ,2008,10 ( 2 ) : 151 - 155.
  • 10Kaltman JR, Ro PS,Zinunerman F, et al. Managed ventricular pacing in pediatric patients and patients with congenital heart disease [ J ]. Am J Cardiol,2008,102 ( 7 ) : 875 - 878.

引证文献3

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部