期刊文献+

有机液、消毒液中相关化合物的GC-MSD分析测定

Analysis of Relative Compounds in Organic and Decon Samples by GC - MSD
下载PDF
导出
摘要   禁止化武公约组织(OPCW)每年都要考核各参试国家的分析能力,我们实验室连续多年参加水平考试,并都取得了很好的成绩.本文报导了第14次OPCW水平考试其中的有机液(O/14)和消毒液(D/14)样品分析.有机液是模拟视查某生产、合成车间,采自1个未作标记包装容器内的样品(编号O/14).…… The fourteenth official OPCW proficiency test organic sample and Decon sample were analyzed by GC-MSD. There were a lot of impurities in the organic sample. Relative compounds were not separated by DBS-MS column, but DB-1701 chromatography column can separate them. Organic sample was extracted by acetonitrile to remove background. The Decon sample was extracted by dichloromethane. All compounds in two samples were determined successfully.
机构地区 防化研究院
出处 《分析测试学报》 CAS CSCD 北大核心 2004年第z1期189-191,共3页 Journal of Instrumental Analysis
关键词 ANALYSIS Relative compounds GC - MSD Analysis Relative compounds GC - MSD
  • 相关文献

参考文献4

  • 1[1]Standard Operating Procedure for the Organisation of OPCW Profiency Tests, 1994, (2).
  • 2[2]Work Intruction (WI) for the Preparation of Test sample foor OPCW Proficiency Tests, 1994, (2).
  • 3[3]Work Instruction(WI) for the Evaluatio of Results of OPCW Proficiency Tests, 1994, (2).
  • 4[4]Report of Fourteenth Official OPCW Proficiency Test 2004.10.

相关作者