期刊文献+

^(99)Tc^m-MIBI门控心肌显像ECTS软件测量左室射血分数 被引量:8

Measurement of left ventricular ejection fraction from gated myocardial perfusion SPECT with ^(99)Tc^m-MIBI by ECTS
下载PDF
导出
摘要 目的 评价99Tcm 甲氧基异丁基异腈 (MIBI)门控心肌显像EmoryCardiacToolboxsoftwere(ECTS)处理软件测量左室射血分数 (LVEF)的临床价值。方法  31例受检者 ,采用静息 多巴酚丁胺负荷一日法 ,行99Tcm MIBI门控心肌断层显像 ,用ECTS软件测量LVEF ,并于 1周内完成平衡法心室显像 ,比较 2种方法测量LVEF的相关性及一致性。结果 ①ECTS软件分析左室功能参数的重复性很好 ,变异系数 (CV)均小于 2 %。②99Tcm MIBI门控心肌显像ECTS软件测定LVEF值与平衡法心室显像测定值相关性良好 (r=0 893,P <0 0 0 1)。③ECTS软件测定的LVEF值 [(5 7 3± 2 93) % ]较心室显像测定值 [(4 6 6± 2 86 ) % ]高 (t=7 76 ,P <0 0 0 1)。结论 99Tcm MIBI门控心肌显像ECTS软件与平衡法心室显像测定LVEF值具有较好的相关性 ,但前者测量值较后者高。99Tcm MIBI门控心肌显像ECTS软件测定LVEF值能否完全取代平衡法心室显像测定 ,尚需进一步研究。 Objective To evaluate clinical value of Emory Cardiac Toolbox processing software (ECTS) measuring the left ventricular ejection fraction (LVEF) from gated myocardial perfusion SPECTS with 99Tcm-methoxyisobutylisonitrile (MIBI). Methods Gated myocardial perfusion SPECTS with 99Tcm-MIBI was performed on 31 cases by using rest-stress with dobutamine protocol in a same day. LVEFs were measured by ECTS. Equilibrium gated blood-pool imaging (EGBI) was also performed within a week. The correlation and conco...
出处 《中华核医学杂志》 CAS CSCD 北大核心 2003年第S1期20-22,共3页 Chinese Journal of Nuclear Medicine
关键词 门控血池显像 MIBI 心脏功能试验 Gated blood-pool imaging MIBI Heart function test
  • 相关文献

参考文献2

二级参考文献24

  • 1周颖,何青,屈婉莹,高文萍,朱明,赵洪山,周清华,高天科.^(99m)Tc-MIBI门控心肌断层显像同时评价心肌灌注和左室功能[J].中华核医学杂志,1995,15(2):87-89. 被引量:20
  • 2曾骏,赵惠扬,王凤琴,黄钢,陈可靖,单越芬.门控^(99m)Tc-MIBI心肌灌注断层显像测量室壁收缩分数的方法学研究[J].中华核医学杂志,1995,15(3):142-144. 被引量:3
  • 3曾骏 王凤琴.门电路心肌灌注断层显像的临床评价[J].国外医学:放射医学核医学分册,1994,18:27-27.
  • 4林景辉 潘中允.心血管系统(Ⅲ)-心血池显像和心室功能测定.临床核医学[M].北京:原子能出版社,1994.210-230.
  • 5[1]Sochor H, Huber K, Probst P, et al. Assessment of myocardial perfusion and wall motion by the new perfusion agent Tc-99m MIBI andSPECT. Eur Heart J, 1988,9: 1-364
  • 6[2]Mochizuki T, Murase K, Fujiwara Y, et al. Assessment of systolicthickening with thallium-201 ECG-gated singlephoton emissioncomputed tomography: a parameter for local left ventricular function.J Nucl Med, 1991.32:1496-1500
  • 7[3]Camargo E, Hironaka FH. giorgi MCP, et al. Amplitude analysis ofstress technetium-99m methoxy isobutylisonitrile images in coronaryartey disease. Eur J Nucl Med, 1992; 19:484-491
  • 8[4]Faber TL, Akers MS. Peshock RM, et al. Three-dimensional motionand perfusion quantification in gated single-photon emission com-puted tomograms. J Nucl Med, 1991,32: 2311-2317
  • 9[5]Clausen M, Henze E, Schmidt A, et al. The contraction fraction(CF)in myocardial studies with Technetium-99m-isonitril (MIBI)-correlations with radionuclide ventriculography and infarct size mea-sured by SPECT. Eur J Nucl Med, 1989,15:661-664
  • 10[6]Pryor DB, Harrel FE,Jr Lee KL, et al. Prognostic indicators from ra-dionuclide angiography in medically treated patients with coronaryartery disease. Am J Cardiol, 1984,53:19-22

共引文献13

同被引文献69

  • 1董军亚,黄体钢,周长钰.左心室射血分数测定方法比较[J].中国心血管杂志,2004,9(5):378-381. 被引量:3
  • 2李河北,王茜,秦淑玲,王玉,黄俐俐.门控心肌显像与心血池显像测量左室射血分数的比较[J].中国医学影像技术,2005,21(7):1082-1084. 被引量:3
  • 3刘敏,郭佑民,杨爱民,邓惠兴,韩克,郭小娟.^(99)Tc^m-MIBI门控心肌灌注断层显像与超声心动测定左室射血分数的对比研究[J].实用放射学杂志,2007,23(4):520-522. 被引量:5
  • 4Sharir T, Germano G, Kavanagh PB, et al. Incremental prognostic value of post-stress left vcntricular ejection fraction and volume by gated myocardial perfusion single photon emission computed tomography. Circulation, 1999, 100: 1035-1042.
  • 5Germano G, Kiat H, Kavanagh PB, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med, 1995, 36 : 2138-2147.
  • 6Faber TL, Cooke CD, Folks RD, et al. Left ventricular function and perfusion from gated SPECT perfusion images: an integrated method. J Nucl Med, 1999, 40: 650-659.
  • 7Ficaro EP, Quaife RA, Kritzman JN, et al. Accuracy and reproducibility of 3D-MSPECT for estimating left ventricular ejection fraction in patients with severe perfusion abnormalities. Circulation, 1999, 100 Suppl 1: 126.
  • 8Nakata T, Katagiri Y, Odawara Y, et al. Two- and three-dimensional assessments of myocardial perfusion and function by using ^99Tc^m sestamibi gated SPECT with a combination of count- and image-based techniques. J Nucl Cardiol, 2000, 7: 623-632.
  • 9He ZX, Cwajg E, Preslar JS, et al. Accuracy of left ventricular ejection fraction determined by gated myocardial perfusion SPECT with ^201Tl and ^99Tc^m-sestamibi: comparison with first-pass radionuclide angiography. J Nucl Cardiol, 1999, 6: 412-417.
  • 10Vourvouri EC, Poldermans D, Bax JJ, et al. Evaluation of left ventricular function and volumes in patients with ischaemic cardiomyopathy: gated single-photon emission computed tomography versus two-dimensional echocardiography. Eur J Nucl Med, 2001, 28: 1610-1615.

引证文献8