期刊文献+

百特OXIM-Ⅱ06 Plus膜肺在小儿外科临床应用(附201例病例报告)

Clinical Application of Baxter OXIM-Ⅱ 06 Plus Infant/Pediatric Membrane Oxygenator--Report of 201 cases
下载PDF
导出
摘要 目的 评价改进型的百特OXIM -Ⅱ 0 6Plus(百特小号 )膜肺的性能和质量。方法 选择百特小号氧合器在体外循环中应用的手术病人 2 0 1例 ,体重 3 8~ 2 0kg,平均 9 14± 3 19kg,体外转流时间为 18~ 2 89min,平均 5 9± 4 3min ,阻断时间为 8~ 12 2min ,平均 3 6± 2 1min。术中监测血气、血红蛋白尿发生率、升降温速率。结果 在体外循环中监测血气都在正常范围内 ,血红蛋白尿发生率低 ( 5 9% ) ,变温效果达到临床要求包括深低温手术 ,体外循环过程平稳 ,且百特小号预充量为 110ml,流量是 2 0L。结论 百特小号氧合气在临床上使用广泛 ,性能较好 ,适合于新生儿到 2 Objective The quality and function of Baxter OXIM-Ⅱ06 Plus(small) membrane oxygenator in clinical application were evaluated.Methods The clinical application of Baxter(small) oxygenator in 201 cases was studied. The patients' weight range from 3.8~20kg〔mean of (9.14±3.19)kg〕; bypass time was from 18~289min〔mean of (59±43)min〕; the aortic cross clamp time ranged from 8~122min〔mean of (36±21)min〕. During the surgical procedure, blood gases, occurrence of hemoglobinurin and blood temperature were monitored. Results The blood gases were within normal rang in all cases during bypass. The occurrence of hemoglobinuria was low(5.9%). It had appropriate thermo-changing effect including deep hypothermia. The procedure was stable during cardiopulmonary bypass. The primary volume was 110ml and the flow rate was 2.0L. Conclusion It was demonstrated that the OXIM-Ⅱ 06 Plus infant/pediatric membrane oxygenator possesses of good function and performs effectively in infants and children under 20kg.
出处 《生物医学工程与临床》 CAS 2001年第4期209-210,共2页 Biomedical Engineering and Clinical Medicine
关键词 膜肺 体外循环 新生儿 membrane oxygenator cardiopulmonary bypass neonate
  • 相关文献

参考文献1

  • 1[3]Blauth C. Influence of oxygenator type on the prevalence and extent of microembolic retinal ischemia during cardiopulmonary bypass. Assessment by digital image analysis[J]. J Thorac Cardiovasc Surg,1990,99(1) :61 - 69

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部