期刊文献+

热电偶和二次仪表现场整体校准系统 被引量:2

Scene Whole Adjust System for Thermocouple and two Degree Instrument
下载PDF
导出
摘要 长期以来,车间仪表人员都是把热电偶和二次仪表分别拆下来,送计量部门检定,检定完毕后再装回原处。热电偶在拆装过程中经常会损坏,并且测试费用非常高,所以有些人常常放弃热电偶的校准工作。热电偶和二次仪表的现场整体校准技术在目前还是一项新技术,结合长安公司兵器工业四五六区域计量站的项目对热电偶、二次仪表整体校准技术进行了一定的研究。介绍了热电偶和二次仪表现场校准系统的基本组成及测试的整个过程;讨论了面向对象的程序设计技术在整体校准系统中的应用方法;最后简要介绍了整体校准系统的主要功能模块和关键技术。 For a long time,the staffs of the workshop dismantle the thermocouple and the two degree instruments respectively and send them to the measure department to be verified,put them back after been calibrated.The thermocouple will often be damaged in the course of disassembly and assembly,and the testing expenses are very high,so some people often give up the calibration work of the thermocouple.It is still a new technology at present to calibrate the thermocouple and two degree instruments at the scenes.Based on the weapon industry 456 area projects,the calibration technology of the thermocouple and the two degree instruments is developed. The basic composition and the whole course tested of thermocouple and two degree instrument scene are introduced the target-oriented procedure in the application method regulating the system designing technique wholly is discussed; the main function moducle and key technology of regulating the system wholly are introduced.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2004年第z3期165-167,178,共4页 Chinese Journal of Scientific Instrument
关键词 热电偶 二次仪表 1560堆桟测温仪 现场整体校准 Thermocouple Two degree instrument 1560 black stack thermometer readout Scene calibrate wholly
  • 相关文献

参考文献3

  • 1[3]国家技术监督局计量司.S型标准热电偶热电势计算手册.北京:中国计量出版社,1993.
  • 2[4]Hart Scientific. 1560 black stack thermometer readout user's guide. American Fork, 2002,5 ~ 110,157 ~ 164.
  • 3[5]Hart Scientific. 9150 portable furnace user's guide.American Fork, 2002,1 ~ 40.

同被引文献7

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部