期刊文献+

SYSMEX尿液常规分析系统的显微镜复查标准探讨 被引量:2

Study of the standard of microscopic reexamination for SYSMEX urine analysis system
下载PDF
导出
摘要 目的探讨SYSMEX尿液干化学分析仪与尿沉渣仪的显微镜复查标准。方法取新鲜尿液进行尿液干化学和尿沉渣分析仪检测,同时使用显微镜尿沉渣计数板定量计数尿沉渣白细胞/红细胞。结果14.16%的尿干化学分析正常样本显微镜复查WBC\RBC结果超出生物参考区间;尿沉渣自动分析与镜检符合率为96.58%,尿干化学分析与镜检符合率为85.84%。结论单一尿干化学分析方法不能替代显微镜复查方法,对该类尿液常规分析有必要做显微镜复查;UF100尿沉渣分析仪测定结果在生物参考区间内的样本无需显微镜复查,超出生物参考区间的样本需显微镜复查。 Objective To discuss the standard of microscopic reexamination for SYSMEX urine dry chemistry analyzer and urine sediment analyzer.Methods Fresh urine is obtained to carry on a urine dry chemistry and sediment analysis,and counting WBC/RBC be performed by using urine sediment count plate.Results Microscopic reviewed WBC\RBC results beyond the biological reference interval in 14.16% of normal dry chemical analysis;there is a 96.58% coincidence rate between urine sediment microscope review and urine sediment ...
出处 《实用医院临床杂志》 2009年第6期86-87,共2页 Practical Journal of Clinical Medicine
关键词 尿干化学分析 尿沉渣自动分析 UF100显微镜 复查 符合率 Urine dry chemistry analysis Urine sediment auto-analysis UF100 Microscopic review Coincidence rate