期刊文献+

水中氰化物的测定及保存研究 被引量:4

Study on the determioation and preservation of cyanide in water
原文传递
导出
摘要 采用模块化流动注射分光光度法对环境水中的氰化物进行了测定,并与国家标准方法进行对照,所测结果经统计学处理,两种方法无显著性差异,精密度、准确度等各项分析指标符合分析标准。同时对不同水体中氰化物保存条件进行了研究,结果表明水样采集后应立即测试,不能立即测试的,加碱保存在4℃不应超过24 h。 The flow injection analysis(FIA) spectrophotometry was employed for the determination of cyanide in water,and the data obtained were compared with those obtained from the standard method of pyridine-barbituric acid spectrophotometry.The results showed no statistically significant differences between these two methods.The precision,accuracy and other indexes accorded with the analysis criteria.Based on the established FIA spectrophotometry,the preservation conditions of cyanide were investigated in water.The...
出处 《分析试验室》 CAS CSCD 北大核心 2010年第5期82-85,共4页 Chinese Journal of Analysis Laboratory
基金 国家地质实验测试中心基本科研业务费(200607CSJ17)项目资助
关键词 氰化物 流动注射分光光度 在线蒸馏模块 水样的保存 Cyanide Flow injection analysis(FIA) spectrophotometry On-line distillation module Water preservation
  • 相关文献

参考文献15

二级参考文献10

  • 1国家环保局.水和废水监测分析方法(第三版)[M].北京:中国环境科学出版社,1989.108,354.
  • 2朱兆海 方肇伦.污水中氰化物的气体扩散分离流动注射分析[J].科学通报,1986,31(10):800-800.
  • 3陈履安(译),水质试验法(修订版),1990年,454页
  • 4Wu X Z,Anal Chem,1989年,61卷,14期,1505页
  • 5国家环保局,水和废水监测分析方法(第3版),1989年,306页
  • 6朱兆海,科学通报,1986年,31卷,10期,800页
  • 7GB/T 7486 1987
  • 8GB/T 7487 1987
  • 9国家环境保护总局,水和废水监测分析方法编委会编,魏复盛.水和废水监测分析方法[M]中国环境科学出版社,2002.
  • 10北京师范大学[M].