期刊文献+

基于金微粒修饰的聚苯胺乙脑疫苗传感器的研究

Studies on a nano Au-modified polyaniline japanese encephalitis vaccine electrode
原文传递
导出
摘要 采用聚合物掺杂法将金微粒固定在聚苯胺膜中,再通过静电吸附固定抗体,制成非标记电流型乙脑疫苗传感器。根据抗原抗体特异性结合形成的免疫复合物使电极敏感膜有效扩散面积减小,电流减少的特性,提出以嵌入金微粒的聚苯胺膜作为电子媒介体,用0.1mol/L磷酸盐缓冲液(pH7.4)为测试底液检测乙脑疫苗效价的新方法。用该传感器检测乙脑疫苗效价,线性范围6.1×10-8~4.1×10-6lgpfu/mL,相关系数r=0.994,检测限为3.0×10-8lgpfu/mL(S/N=3)。该传感器可用于生物制品的测定。 A label-free amperometric Japanese encephalitis vaccine electrode was prepared by immobilization of nano-Au in polyaniline film by electrochemical doping method,and antibody of Japanese encephalitis vaccine was immobilized on nano-Au by attraction. The effective diffusion section area of electron attained to the electrode decreased with the increase of the singularity reaction immunocompound of antibody and antigen,so Japanese encephalitis vaccine could be determined by utilizing nano Au-modified polyanilin...
出处 《分析试验室》 CAS CSCD 北大核心 2010年第7期69-71,共3页 Chinese Journal of Analysis Laboratory
基金 遵义医学院硕士科研启动基金(2008F-303)项目资助
关键词 金微粒 聚苯胺膜 聚合物掺杂法 乙脑疫苗 非标记电流型传感器 Polyaniline film Nano-Au Japanese encephalitis vaccine Label-free amperometric immunosensor
  • 相关文献

参考文献3

  • 1张凌燕,黎雪莲,袁若.高灵敏电流型免疫传感器对流行性乙脑疫苗的测定(英文)[J].西南大学学报(自然科学版),2008,30(3):51-55. 被引量:1
  • 2Tang,D.-P.,Yuan,R.,Chai,Y.-Q. Journal of Physical Chemistry B . 2006
  • 3Mikhaylova A A,Molodkina E B,Khazova O A,et al.Electrocatalytic and adsorption properties of platinum microparticle electrodeposited into polyaniline films[].Journal of Electroanalytical Chemistry.2001

二级参考文献8

  • 1唐点平,袁若,柴雅琴,刘颜,钟霞,戴建远.基于聚乙烯醇缩丁醛和金溶胶修饰的电位型铂金免疫传感器的研制[J].西南师范大学学报(自然科学版),2004,29(5):843-847. 被引量:5
  • 2朱强,袁若,柴雅琴,黎雪莲,王娜.以纳米金为介质的无标记电流型甲胎蛋白免疫传感器的研究[J].西南师范大学学报(自然科学版),2006,31(1):81-85. 被引量:11
  • 3[1]Wang B Z,He C M.The Basis of Biological Products[M].Gansu:Peoples Publishing House,1986:25.
  • 4[2]S Babacan S,Pivarnik P,Letcher S,Rand A G.Evaluation of Antibody Immobilization Methods for Piezoelectric Biosensor Application[J].Biosensors and Bioelectronics,2000,15:615-621.
  • 5[5]Santandreu M,Cespedes F,Alegret S,Esteve Martínez-F.Amperometric Immunosensors Based on Rigid Conducting Immunocomposites[J].Anal Chem,1997,69:2080-2085.
  • 6[6]Wendzinski F,Grundig B,Rennerberg R,et al.Highly Sensitive Determination of Hydrogen Peroxide and Peroxidase with Tetrathiafulvalene-Based Electrodes and the Application in Immunosensing.Biosens[J].Bioelectron,1997,12:43-52.
  • 7[7]Kalab T,Skladal P.A Disposable Amperometric Immunosensor for 2,4-Dichlorophenoxyacetic Acid[J].Anal Chim Acta,1995,304:361-368.
  • 8[8]Yuan R,Zhang L Y,Li Q F,et al.A Label-Free Amperometric Immunosenor Based on Multi-Layer Assembly of Polymerized o-Phenylenediamine and Gold nanoparticles for Determination of Japanese B Encephalitis Vaccine[J].Anal Chim Acta,2005,531:1-5.