依据JJG1059-1999对直读光谱仪碳元素检出限的不确定度进行了评估,分析了影响测量不确定度的各个因素,计算得出直读光谱仪碳元素检出限的扩展不确定度为:U=0.11%,k=2。
Based on JJG1059-1999,the uncertainty of detection limit of carbon with direct reading spectrometer was assessed.A,the same time,the various factors which effect the uncertainty of measurement were analyzed. Then,the expanded uncertainty was obtained by calculation:u = 0.0011%,k = 2.
出处
《黑龙江环境通报》
2010年第4期48-50,共3页
Heilongjiang Environmental Journal
关键词
直读光谱仪
碳
检出限
不确定度
Direct reading spectrometer
Carbon
Limit of detection
Uncertainty