摘要 目的:通过介绍CTVision系统的结构和技术特点,探讨其在图像引导放射治疗中的应用方式。方法:将CTVision系统与目前常用的三维IGRT系统,特别是kVCBCT系统进行图像质量对比,得出其临床技术特点及应用方式,并以CTVision在线校位应用方式为例,分析临床操作流程及治疗精度。结果:用CTVision进行IGRT在线校位,可以减少肿瘤分次治疗间摆位误差和形变误差,几何精度可达1mm。结论:CTVision图像引导放射治疗应用方式简单可行,能保证治疗靶区的准确照射,可进一步提高肿瘤放射治疗的精度。 Objective:By introducing the structure and technical characteristics of CTVision system,we discusse its application modes in image guided radiotherapy.Methods:Compare the image quality of CTVision with other 3D IGRT system,especially kV CBCT,we get its clinical technical characteristics and application modes,then,analyse its clinical operation process and treat precision by its online correction technology.Results:The online correction technology of CTVision can reduce the setup error and deformation error ...
出处 《中国医学物理学杂志》 CSCD 2011年第1期2382-2385,共4页 Chinese Journal of Medical Physics
基金 广东省科技计划项目(No.2010B031100013)
关键词 CTVision 图像引导 在线校位 CTVision image guided online correction
  • 相关文献

参考文献2

二级参考文献41

  • 1.调强适形放射治疗[A].胡逸民.肿瘤放射物理学[C].北京:原子能出版社,1999.538-612.
  • 2Hanley J,Debois MM,Mah D,et al.Deep inspiration breath-hold technique for lung tumors:the potential value of target immobilization and reduced lung density in dose escalation.Int J Radiat Oncol Biol Phys,1999,45:603-611.
  • 3Ramsey CR,Scaperoth D,Arwcod D,et al.Clinical efficacy of respiratory gated conformal radiation therapy.Med Dosim,1999,24:115-119.
  • 4Shirato H,Shimizu S,Kitamura K,et al.Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.Int J Radiat Oncol Biol Phys,2000,48:435-442.
  • 5Keall PJ,Chen GTY,Joshi S,et al.Time-the fourth dimension in radiotherapy (ASTRO panel discussion).Int J Radiat Oncol Biol Phys,2003,57:S8-S9.
  • 6Yu CX,Jaffray DA,Wong JW.The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation.Phys Med Biol,1998,43:91-104.
  • 7de Neve W,Heuvel F,De Beukeleer M,et al.Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch.Radiother Oncol,1992,24:45-54.
  • 8Hu YM,DM JR,Hu B.The precision and accuracy analysis of target position in X-ray stereotactic radiotherapy for extracranial lesions.In:Internation symposium of stereotactic radiosurgery (SRS/SRT).Taejon:Chungnam National University Hospital,1997.
  • 9Yan D,Vicini F,Wong J,et al.Adaptive radiation therapy.Phys Med Biol,1997,42:4123-4132.
  • 10Birkner M,Yan D,Alber M,et al.Adapting inverse planning to patient and organ geometrical variation algorithm and implementation.Med Phys,2003,30:2822-2831.

共引文献185

同被引文献44

引证文献5

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部