摘要
目的:分析影像引导放疗中常用的模拟定位螺旋CT(MSCT)、加速器机载锥形束CT(CBCT)和螺旋断层治疗机MV螺旋CT(MVCT)的影像质量。方法:CatPhan600模体测量三种成像设备的CT值均匀性和线性、空间分辨率、低对比度分辨率、3D影像的几何准确性和金属伪影大小。结果:MSCT、CBCT和MVCT的CT值与物质密度均呈线性关系;均匀模体中的CT均匀性分别为1.07%,1.40%,39.81%;空间分辨率分别为8 lp/cm,7 lp/cm,4 lp/cm;MSCT可显示低对比度为1%的3 mm圆孔,CBCT只能略微分辨圆孔,而MVCT则无法显示圆孔,几何准确性均良好;MVCT的金属伪影较小。结论:CBCT和MVCT的成像质量均不如MSCT,但能满足患者治疗体位验证和监测的需要。
Objective: To evaluate the image qualities of Multi-slice CT(MSCT),Cone-beam CT(CBCT),Megavoltage CT(MVCT) commonly used in image-guided radiotherapy.Methods: The CT uniformity and linearity,spatial resolution,low contrast resolution and geometrical accuracy,metal artifacts in three CT modes were evaluated with Catphan600 phantom.Results: CT values depended linearly on the physical densities of the samples for all three CT modes,but maximum noise was found in MVCT.The CT uniformity for MSCT,CBCT,and MVCT wa...
出处
《中国医学物理学杂志》
CSCD
2011年第2期2523-2528,共6页
Chinese Journal of Medical Physics