6MV医用直线电子加速器束流中心方向调整方法的研究 被引量:1

Investigation of Beam Alignment of 6 MV Medical Linear Accelerator
下载PDF
导出
摘要 目的:在对6 MV低能电子直线加速器的加速管安装调试过程中,找出一种简单易行的束流调整方法,使束流调整后辐射野的均整度和方形X-辐射野的对称性指标符合国家标准GB15213-94《医用电子加速器性能及试验方法》的要求。方法:经过分析发现束流对称性(SYM)误差产生的原因分为两部分:束流的方向相对于辐射头轴线发生角度Δθ偏离和束流入射位置相对于辐射头轴线发生ΔX偏离,通过试验得到Δθ和ΔX对SYM数值变化的曲线,通过修正加速管连接法兰盘的平行度和同轴度对束流进行调整,并用水箱测量一定深度处辐射野内剂量的均匀性对调整结果进行检查和分析。结果和结论:使用5支加速管进行试验,束流调整前的对称性值在1.0%至2.9%之间,束流调整后的对称性值在0.8%~1.0%之间。达到了束流调整的目地,并且本方法在多台电子直线加速器上使用,结果证明了该方法的有效性和结果的精确性。 Objective: During installation and adjusting of accelerator tube for 6 MV medical linear accelerator(LINAC),an easy method was invented judged by parallelism and the coaxiality evaluation of the accelerating waveguide connection plate.Methods:The beam direction deviated the accelerator main axis Δθ and the deviation of the beam center ΔX were adjusted.The dose profiles got by the water tank scanner system were used to verify the revising by their flatness and symmetry.Results and Conclusions: The method has been used to adjusting the several accelerator tubes,and the results showed its effective and accuracy enough.
出处 《中国医学物理学杂志》 CSCD 2011年第4期2707-2711,共5页 Chinese Journal of Medical Physics
基金 973项目课题(No.2011CB707701) 北京市教委的共建项目(No.D100010501)
关键词 医用电子直线加速器 束流调整 靶点 medical electron linear accelerator beam alignment target
  • 相关文献

参考文献2

二级参考文献11

  • 1Landry DJ, Anderam DW. Measurement of Accelerator Bremsstralung Spectra With a High Efficiency Ge Detector[J]. Medical Physics, 1991,18: 572-532.
  • 2Faddegon BA, Ross CK, Rogers DWO. Forwarddirected Bremsstralung of 10 to 30 MeV Electrons Incident on Thick Targets of Al and Pb[J]. Medical Physics, 1990, 17:773-785.
  • 3Nath R, Sehulz RJ. Photoactivation Ratio for Specification of High-energy X-ray Quality: Part Ⅰ,15-40 MV X-ray[J]. Medical Physics, 1977,4:36-41.
  • 4Prado KL, Anderson DW. Photonudear Activation Ratios in Fluorire Comapounds:an Index of Brenasstranhung Quality [J]. Medical Physics,1984, 11:474-479.
  • 5Huang PH, Kase KR, Bjarngard BE. Sepctral Characterization of 4 MV Bremsstralung by Attenuation Analysis[J]. Medical Physics, 1981, 8:368-374.
  • 6Huang PH, Kase KR, Bjamgard BE. Simulation Studies of 4 MV X-ray Spectral Reconstruction by Numerical Analysis of Transmission Data [J].Medical Physics, 1982, 9:707-710.
  • 7Tominaga S. The Estimation of X-ray Spectral Distributions From Attenuation Data by Means of Iterative Computation [J]. Nucl Instrona Methods,1981,192: 88-92.
  • 8Piermattei A. A Study of Quality of Bremsstahlung Spectra Reconstructed From Transmission Measurements[J]. Medical Physics, 1990, 17:227-233.
  • 9Nordell B, Brahme A. Phys. Med. Biol., 1984, 29(7): 797-810.
  • 10施将君,钱民权.电子束参量对轫致辐射照射量角分布的影响[J].强激光与粒子束,1992,4(1):119-126. 被引量:3

共引文献23

同被引文献19

引证文献1

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部