期刊文献+

低温石英天平在材料放气污染特性测试中的应用 被引量:3

APPLICATION OF LOW TEMPERATURE QCM TO OUTGASSING CONTAMINATION CHARACTERISTICS DETECTION OF SPACECRAFT MATERIALS
下载PDF
导出
摘要 研制了适用于ASTME 1559航天器材料放气污染特性测试标准的低温石英晶体微量天平,天平的谐振频率可以为5MHz、10MHz和20MHz,能够在82K的低温下工作。文章对天平的输出频率稳定性、质量灵敏度和低温凝结效应进行了测量,结果表明该天平在各个方面均能满足ASTM E 1559标准测试要求。 A series of low temperature quartz crystal microbalance (QCM) is developed to match ASTM E1559,standard test method for contamination outgassing characteristics of spacecraft materials,which resonant frequency of QCM can be 5 MHz,10 MHz or 20 MHz.The lowest working temperature is 82 K.Many measurements are taken to determine the stability of output frequency,mass sensitivity and condensation effect on cryogenic surface.The results demonstrate that the QCMs meet the requirement of ASTM E 1559 well.
出处 《航天器环境工程》 2005年第5期300-303,共4页 Spacecraft Environment Engineering
关键词 石英晶体微量天平 材料放气 污染 quartz crystal microbalance material outgassing contamination.
  • 相关文献

参考文献1

二级参考文献3

  • 1[2]ASTM Designation:E595-90.Standard Test Method for Mass Loss and Collected volatile Condensable Materials from Outgassing.Published April 1990
  • 2[3]ESA PSS-01-702 Issue 1.A Thermal Vacuum Test for the Screening of Space Materials.March 1983
  • 3[6]AD-A267 202:Solar Absorptance of Optical Surfaces Contaminated with Spacecraft Material Outgassing Products.Jul,93