期刊文献+

原子吸收火焰法测定真空采血管中NaF含量的不确定度评估

The Uncertainty Evaluation of NaF Content in Evacuated Receptacle by Flame Atomic Absorption Spectrometry
下载PDF
导出
摘要 建立原子吸收法火焰法(FAAS)测定真空采血管中NaF含量的不确定度评估方法。根据测量过程,全面鉴别不确定度来源,计算各分量的不确定度,建立数学模型,最后计算出合成标准不确定度和扩展不确定度。真空采血管中NaF含量测定的结果可表示为2.15±0.052 mg/L(k=2),合成标准不确定度为0.026 mg/L和扩展不确定度为0.052 mg/L。本方法适用于原子吸收法火焰法测定真空采血管中NaF含量的不确定度评估,对合理地赋予被测定结果的分散性具有重要意义。 To establish a method for the evaluation of the uncertainty in the NaF content determination of evacuated receptacle by FAAS.According to the measurement process,each component of uncertainty was calculated by analyzing the various variable sources of uncertainty generally,and a mathematical medel was established.The combined standard uncertainty and the expanded uncertainty were then finally obtained.The expression of the results of the NaF content determination of evacuated receptacle was 2.15 ±0.052 mg/L (k =2),the combined standard uncertaintywas 0.026 mg/L,the expanded uncertainty was 0.052 mg/L.This method is suitable for evaluating of the NaF content determination of evacuated receptacle by FAAS,and it is important meaning to give the dispersion of the determined results.
出处 《生物医学工程研究》 2013年第4期249-252,共4页 Journal Of Biomedical Engineering Research
关键词 原子吸收 火焰法 采血管 NAF 不确定度 Atomic absorption spectrometry Flame Blood specimen collection containers NaF Uncertainty in measurement
  • 相关文献

参考文献2

  • 1中国实验室国家认可委员会.实验室认可与管理基础知识[M]{H}北京:中国计量出版社,2003.
  • 2国家质量技术监督局计量司.测量不确定度评定与表示指南[M]{H}北京:中国计量出版社,2000.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部