期刊文献+

Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients

Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients
下载PDF
导出
摘要 AIM:To evaluate the measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.·METHODS:A hundred and one eyes of 55 keratoconus patients were enrolled in this study.The mean age was26.2 ±8.9 years.The inclusion criteria were keratoconus stage I to III according to the Amsler-Krumeich keratoconus classification.All the measurements were done by the same operator,under the mesopic light condition and repeated with three different optical methods;Visante,Orbscan and Pentacam.The evaluated anterior segment parameters were anterior chamber depth(ACD),central and thinnest corneal thickness(CCT and TCT) and pupil diameter(PD).·RESULTS:The mean CCT measured by Visante,Orbscan and Pentacam were as follows:462.0 ±48.1μm,463.9±60.9μm,476.5±45.3μm,respectively(P =0.873).The mean ACD values were 3.34±0.33mm,3.26±0.33mm,3.49±0.40mm,respectively(P=0.118).The mean PD measurements were 5.11 ±1.14mm,4.80 ±0.85mm,3.80 ±1.38mm,respectively(P 【0.001).The mean TCT measurements of Visante,Orbscan and Pentacam were 437.9 ±48.2μm,447.6±60.6μm and 459.9±44.0μm,respectively(P =0.214).The Visante and Orbscan measured CCT similarly,while Pentacam measured CCT thicker than the other two.The Visante measured TCT thinner than the other two devices.In ACD measurements,Orbscan was the one giving the lowest values.PD was measured differently by the devices.·CONCLUSION:Although TCT,CCT and ACD measurements acquired by Visante,Orbscan and Pentacam in keratoconus patients are similar,PDmeasurements show large differences among the devices. AIM:To evaluate the measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.·METHODS:A hundred and one eyes of 55 keratoconus patients were enrolled in this study.The mean age was26.2 ±8.9 years.The inclusion criteria were keratoconus stage I to III according to the Amsler-Krumeich keratoconus classification.All the measurements were done by the same operator,under the mesopic light condition and repeated with three different optical methods;Visante,Orbscan and Pentacam.The evaluated anterior segment parameters were anterior chamber depth(ACD),central and thinnest corneal thickness(CCT and TCT) and pupil diameter(PD).·RESULTS:The mean CCT measured by Visante,Orbscan and Pentacam were as follows:462.0 ±48.1μm,463.9±60.9μm,476.5±45.3μm,respectively(P =0.873).The mean ACD values were 3.34±0.33mm,3.26±0.33mm,3.49±0.40mm,respectively(P=0.118).The mean PD measurements were 5.11 ±1.14mm,4.80 ±0.85mm,3.80 ±1.38mm,respectively(P <0.001).The mean TCT measurements of Visante,Orbscan and Pentacam were 437.9 ±48.2μm,447.6±60.6μm and 459.9±44.0μm,respectively(P =0.214).The Visante and Orbscan measured CCT similarly,while Pentacam measured CCT thicker than the other two.The Visante measured TCT thinner than the other two devices.In ACD measurements,Orbscan was the one giving the lowest values.PD was measured differently by the devices.·CONCLUSION:Although TCT,CCT and ACD measurements acquired by Visante,Orbscan and Pentacam in keratoconus patients are similar,PDmeasurements show large differences among the devices.
出处 《International Journal of Ophthalmology(English edition)》 SCIE CAS 2013年第4期521-525,共5页 国际眼科杂志(英文版)
关键词 Visante ORBSCAN PENTACAM corneal pachymetry anterior chamber depth pupil diameter Visante Orbscan Pentacam corneal pachymetry anterior chamber depth pupil diameter
  • 相关文献

参考文献22

  • 1Vishal Jhanji,Bingzhi Yang,Marco Yu,Cong Ye,Christopher KS Leung.Corneal thickness and elevation measurements using swept‐source optical coherence tomography and slit scanning topography in normal and keratoconic eyes[J]. Clin Experiment Ophthalmol . 2013 (8)
  • 2Dinc, Umut Asli,Gorgun, Ebru,Oncel, Banu,Yenerel, Melda Nursal,Alimgil, Levent.Assessment of Anterior Chamber Depth Using Visante Optical Coherence Tomography, Slitlamp Optical Coherence Tomography, IOL Master, Pentacam and Orbscan IIz[J]. Ophthalmologica . 2010 (6)
  • 3Charles R. Edmonds,Shu-Fen Wung,Bart Pemberton,Steven Surrett.Comparison of Anterior Chamber Depth of Normal and Keratoconus Eyes Using Scheimpflug Photography[J]. Eye & Contact Lens: Science & Clinical Practice . 2009 (3)
  • 4Paul S. Zhao,Tien Yin Wong,Wan-Ling Wong,Seang-Mei Saw,Tin Aung.Comparison of Central Corneal Thickness Measurements by Visante Anterior Segment Optical Coherence Tomography With Ultrasound Pachymetry[J]. American Journal of Ophthalmology . 2007 (6)
  • 5Emmy Y.M. Li,Shaheeda Mohamed,Christopher K.S. Leung,Srinivas K. Rao,Arthur C.K. Cheng,Carol Y.L. Cheung,Dennis S.C. Lam.Agreement among 3 Methods to Measure Corneal Thickness: Ultrasound Pachymetry, Orbscan II, and Visante Anterior Segment Optical Coherence Tomography[J]. Ophthalmology . 2007 (10)
  • 6Miyuki Fujioka,Makoto Nakamura,Yasuko Tatsumi,Azusa Kusuhara,Hidetaka Maeda,Akira Negi.Comparison of Pentacam Scheimpflug Camera with Ultrasound Pachymetry and Noncontact Specular Microscopy in Measuring Central Corneal Thickness[J]. Current Eye Research . 2007 (2)
  • 7Susannah Quisling,Stacy Sjoberg,Bridget Zimmerman,Kenneth Goins,John Sutphin.Comparison of Pentacam and Orbscan IIz on Posterior Curvature Topography Measurements in Keratoconus Eyes[J]. Ophthalmology . 2006 (9)
  • 8Nicolás Charles,Martin Charles,J. Oscar Croxatto,Daniel E. Charles,Diego Wertheimer.Surface and Orbscan II slit-scanning elevation topography in circumscribed posterior keratoconus[J]. Journal of Cataract & Refractive Surgery . 2005 (3)
  • 9Gerard Cairns,Charles N.J. McGhee.Orbscan computerized topography: Attributes, applications, and limitations[J]. Journal of Cataract & Refractive Surgery . 2005 (1)
  • 10Georges Baikoff,Eric Lutun,Caroline Ferraz,Jay Wei.Static and dynamic analysis of the anterior segment with optical coherence tomography[J]. Journal of Cataract & Refractive Surgery . 2004 (9)

共引文献3