期刊文献+

DATA PROCESSING ON LINEARIZATION OF HOT-WIRE ANEMOMETER

DATA PROCESSING ON LINEARIZATION OF HOT-WIRE ANEMOMETER
下载PDF
导出
摘要 A method of data processing to determine the coefficients of linearization equations for 1050 anemometer (produced by Thermo-Systems Inc. -TSI, USA) with the sensors made of domestic hot wire using the program preferred in this Paper is described. By calculation and test, it is indicated that the error resulting from this method is about 0. 5% of the full scale and less than TSl's. By using this method we can set up the calibration curve according to the measurement range and the diameter of the hot wire at a certain accuracy.
出处 《Transactions of Nanjing University of Aeronautics and Astronautics》 EI 1994年第2期224-228,共5页 南京航空航天大学学报(英文版)
关键词 measurements data PROCESSING LINEARIZATION aero dynamic test HOT-WIRE ANEMOMETERS measurements data processing linearization aero dynamic test hot-wire anemometers
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部