摘要
利用CS3000血细胞分离机对20例献血员实施YRBC单采。结果显示:①献血员血球压积与YRBC溢出时血浆流率呈负相关(Y=67.22-1.111,r=-0.8069,P<0.01),根据献血员Hct来选择血浆流率。可避免非年轻红细胞溢出。②CS3000血细胞分离机可用来获得高质量YRBC供临床应用。
Using CS3000 cell separator enforced young red blood cell(YRBC) apheresis was a negative relation between the packed cell volume(PCV) of the blood donors and the plasma flow rate when YRBC overflowed from the sparator chamber(Y=67. 22-1. 11x,r=-0. 8069, P<0.01). This demonstrates that auording to PCV of the blood donors, the plasma flow rate can be estimated so as to avoid non-YRBC overflowing from separator chamber, and CS3000cell separator can be used in obtaining YRBC of high quality for clinical application.
出处
《河南科技大学学报(医学版)》
1994年第4期212-215,共4页
Journal of Henan University of Science & Technology:Medical Science
关键词
年轻红细胞单采
丙酮酸激酶
葡萄糖-6磷酸脱氢酶
红细胞变形能力
young red blood cell apheresis
pyruvate kinase (PK)
glucose-6-phosphate depydrogenase (G6PD)
red blood cell deformability