期刊文献+

FAAS法测定人尿中锂的研究 被引量:1

STUDY OF DETERMINATION OF LITHIUM IN URINE BY FLAME ATOM ABSORPTION SPECTROMETRY
下载PDF
导出
摘要 本文试验了40种无机离子和20种有机物对锂吸收信号的影响.在给定浓度范围内,无机离子对尿中锂的测定无干扰(α=0.05).有机物的干扰采用NNO_3-H_2O_2湿法消解样品来消除.本文获得的特征浓度(1%吸光度)为0. 04μg/ml,7次测定的R.S.D为5.9%. In this paper, influences of 40 inorganic ions and 20 organic substances on absorption of lithium are studied, showing ex-perimently that these inorganic ions do not influence the determination of lithium in given range of concentration. The influ-ences of organic substances in urine are eliminated by wet-digestion with HNO_3--H_2O_2. The characteristic concentration of themeans (1% absortion) is 0.04 ug/ml and R S D of 7 measurements to digestion urine samples is 5.9%.
出处 《四川师范大学学报(自然科学版)》 CAS CSCD 1992年第5期88-92,共5页 Journal of Sichuan Normal University(Natural Science)
关键词 测定 尿 火焰原子吸收光谱 lithium determination urine FAAS
  • 相关文献

参考文献1

  • 1Y. Feng,K. Bencze,Ch. Pelikan. Lithiumbestimmung mit Matrixmodifier KH2PO4 in Triton X-100 im biologischen Material[J] 1987,Fresenius’ Zeitschrift für Analytische Chemie(5):595~599

同被引文献2

引证文献1

相关作者

内容加载中请稍等...