期刊文献+

气相色谱法快速测定过氧化苯甲酸叔丁酯(CP-02)中微量水分

Fast Gas Chromatography Analysis of Trace Water in Tert Butylperoxy Benzoate (CP-02)
下载PDF
导出
摘要 以GDX-102填充柱气相色谱分析热不稳定的过氧化苯甲酸叔丁酯中ppm级微量水份,柱温、汽化室和检测室最佳温度值分别为110℃、120℃和130℃。水和过氧化苯甲酸叔丁酯保留时间相差30min以上,因此可以连续进样测定水份。6次测定水份的变异系数为2.7%,检测下限浓度为2.7ppm。 Trace water in quantity of ppm in the heat-instable tert butylperoxy benzoate wasanalysed by GDX-102 filling-column gas chromatography with the best column-temperature,evaperating-temperature and examing-temperature 110℃, 120℃ and 130℃ respectly. Thedifference of r. t. between water and tert butylperoxy benzoate is more than 30 minutes, sothe samples can be injected continuosly to analyse trace water. The variation coeffiecient in 6examinations is 2.7%. The low limit of sensitivity in the examinations is 2.7 ppm.
机构地区 兰州助剂厂
出处 《西北师范大学学报(自然科学版)》 CAS 1992年第3期47-50,共4页 Journal of Northwest Normal University(Natural Science)
关键词 微量水份 过氧化苯甲酸叔丁酯 气相色谱 trace water tert butylperoxy benzoate gas chromatography
  • 相关文献

参考文献1

  • 1胡依理.费休定水法的改进[J]化学世界,1957(09).

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部