期刊文献+

电化学法检测植物油中维生素E的不确定度分析 被引量:3

Uncertainty Evaluation of Determination of Vitamin E in Vegetable Oil Using Electrochemical Detection
下载PDF
导出
摘要 利用自主研制的聚吡咯修饰电极测定植物油中VE的含量,通过对检测过程中称量、玻璃量器、标准曲线、电极以及回收率等影响不确定度的因素进行分析,计算不确定度。结果表明,VE检测结果为258.18 mg/kg的油样,其相对标准不确定度为7.36mg/kg,相对扩展不确定度为5.7%,不确定度主要由a类不确定度与玻璃量器的不确定度引起。 We used the independently developed electrode modified by polypyrrole , to determine the content of vitamin E in vegetable oil. The uncertainty was calculated by analysing the affecting factors in the detection process, such as weighing, glass container, standard curve, electrodes, recovery, and so on. The results indicated that, the oil sample whose vitamin E test results was 258.18 mg/kg, was 7.36 mg/kg for the relative standard uncertainty and 5.7%for the relative expanded uncertainty , and the uncertainty was mainly caused by type A uncertainty and uncertainty of the glass container.
出处 《食品研究与开发》 CAS 北大核心 2013年第24期189-192,共4页 Food Research and Development
基金 公益性行业科研专项(200910034) 国家自然基金(20876165)
关键词 电化学法 VE 不确定度 electrochemical detection Vitamin E uncertainty evaluation